WARNING: NOT EXPLOSION PROOF! Using this instrument in an area where devices with intrinsically safe
classification is recommended, is prohibited!
WARNING: Detachable main power supply cord with inadequate ratings should not be used.
WARNING: It is the responsibility of the operator to ensure that pressure in circuits connected to the UUT
will never exceed their safe working pressure. In the event that the maximum safe working pressure of a
circuit / system to be tested is lower than the maximum obtainable output pressure of the UUT, the
operator who will perform the tests must be instructed in advance regarding the maximum safe working
pressure of the external circuit.
Depending on the volume and pressure in the UUT, the risks for humans and material can be very high. For
that reason the end user must ensure that maximum pressure levels for each circuit are established as set
forth by law. (European directive 2014/68/EU or similar guidelines/obligations are specified and applicable
in the country of use.)
Because of the potential risk for human beings in the vicinity of systems/circuits under pressure, the end
user must ensure that risks are kept to the minimum by fencing, warning, forbidding passage or access to
the area for as long as the risk exists.
Do not connect the voltages higher than specified in the specifications to the instrument. If higher voltages
are connected, the interface control may be damaged.
This instrument is equipped with an overpressure protection, which protects the pressure sensor against
overpressure. As soon as the pressure on the pressure sensor increases to 1 bar above the measuring range
specified for the instrument, the supply will be cut off to protect the pressure sensor from damage.