TQC LTD7015 Manual Download Page 1

1 |

TQC ULTRASONIC THICKNESS GAUGE BASIC 

LD7015 

1

 

PRODUCT DESCRIPTION 

Ultrasonic thickness gauge specifically designed to measure the thickness of 
metallic and non-metallic materials e.g. aluminium, titanium, plastics, ceramics, 
glass and plastics. It can also be used to monitor all types of pipes and pressure 
vessels for loss of thickness due to corrosion or erosion. The gauge is easy to use 
and will give accurate readings to an accuracy of 1%. This unit is not suitable for 
cast iron due to its big crystalloid composition. 

1.1

 

Specifications 

Measuring method:  

Ultrasonic pulse echo 

Measuring frequency: 

5MHz  

Measuring range: 

1.20 - 220 mm (steel); (deviation possible with use of other materials) 

Measuring range (tubes): 

20x3mm (steel) 

Nauwkeurigheid:  

+/- (1%H 

+

 0.1) mm 

Display:   

LCD  

Resolution:

0.1mm

Operation temperature:  

0 ºC ~ 40 ºC 

Batteries:  

3x AAA alkaline batteries (total 5V) 

Size:  

70x135x38mm  

1.2

 

Details 

- Receiving plug 
- Transmitting plug 

- Ultrasonic probe 

- LCD screen 

- Calibration block 

Manual LD7016: page 6-13

Summary of Contents for LTD7015

Page 1: ...will give accurate readings to an accuracy of 1 This unit is not suitable for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring freq...

Page 2: ...ng temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored The transducer is made of propyle...

Page 3: ...eted After calibration the sound velocity will remain the same as previous selected A calibration should be made the every time the probe or batteries are replaced This is very important to make sure...

Page 4: ...When the icon flashes please replace the batteries before taking any new measurements Turn the instrument off by pressing the ON OFF button Open the battery lid and replace them with 3 pcs of AAA alk...

Page 5: ...or otherwise is correct we have no control over either the quality or condition of the product or the many factors affecting the use and application of the product Therefore unless we specifically agr...

Page 6: ...e for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring frequency 5MHz 2 5MHz Measuring range Accuracy Sound velocity Display Resolut...

Page 7: ...utor for a replacement probe Working temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored...

Page 8: ...small amount of coupling gel on the sample block and then place the probe on the sample block When the display shows 4 0mm the calibration is completed After calibration the sound velocity will remain...

Page 9: ...ss MODE again to store After selecting your memory unit you can start measuring again While measuring the memory unit will flash and the new measurement will delete the stored value When your measurem...

Page 10: ...the axis of the pipe to be measured For small pipes two measurements should be performed one with the axis of the pipe perpendicular the other one parallel The smallest of the 2 values displayed shou...

Page 11: ...calibrating the gauge with a reference block However this table is just for reference sometimes the sound velocity will be different caused by different physical chemical factors The sound velocity of...

Page 12: ...ve recess surface incorrect calibration sample blocks or the inner flaw of materials 8 Coupling gel Coupling gel causes the high frequency ultrasonic waves transmitting between the probe and material...

Page 13: ...t in use We recommend annual calibration 9 DISCLAIMER The right of technical modifications is reserved The information given in this manual is not intended to be exhaustive and any person using the pr...