2. Simulating pH Values
1. Obtain the correct pH / mV output cable for the instrument under test.
2. Connect one end of the pH / mV output cable to the
WP-Sim
and the
other end to the instrument under test, as per the example diagram
below…
3. Switch the
WP-Sim
Function switch to ON. Switch the instrument
under test on and select pH mode. See the instrument’s handbook if
necessary.
4. Use the pH / mV selector to output any of the 5 simulated pH values.
The readout of the instrument under test should approximately
correspond to the simulated pH value selected on the
WP-Sim
.
Variations are possible due to the current temperature and calibration
settings of the instrument under test.
Please note that the simulated values are for 25
O
C.