10. When the reading has stabilised, adjust the
SPAN
control until the display
shows the value of the standard.
11. Rinse the Conductivity sensor in distilled water and blot dry.
12. The
mini
CHEM
-Cond
is now calibrated and ready for Conductivity
measurements.
4.2 Calibration Notes
1. The Zero calibration is quite stable long term, and only needs to be performed
monthly as a routine check. In applications where the sensor can become dirty
or coated with oils etc., a Zero calibration may need to be done weekly.
2. A Span calibration should be performed at least weekly. Of course, more
frequent calibration will result in greater confidence in results.