7
1.10.3 TDS
Ranges
Resolution
Accuracy
k=0.1 Sensor
0 to 1.000 ppM
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0.001 ppM
0.01 ppM
0.1 ppM
1 ppM
±0.5% of full scale of
selected range at 25
O
C
k=1.0 Sensor
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0.01 ppM
0.1 ppM
1 ppM
0.01 ppK
±0.5% of full scale of
selected range at 25
O
C
k=10 Sensor
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK
0.1 ppM
1 ppM
0.01 ppK
0.1 ppK
±0.5% of full scale of
selected range at 25
O
C
Note : Ranges are automatically selected. Exact auto-ranging points and full scales are subject to
sensor performance.
Sensor Type
...........................................Ryton
®
plastic body with two platinised platinum plates
with in-built ATC.
Temperature Compensation
.................Automatic, 0 to 100
O
C
(standard Conductivity sensor is limited to 60
O
C)
Calibration
............................................Automatic zero and span calibration.
Sensor Span Range
...............................±25% of nominal k factor.
1.10.4 pH
Range
Resolution
Accuracy
0 to 14.00 pH
0.01 pH
±0.01 pH
Sensor Type
...........................................Glass bulb pH sensor, combination or half cell.
Input Impedance
...................................>3 x 10
12
Ohms
Temperature Compensation
.................Automatic, 0 to 100
O
C
Calibration
............................................Automatic asymmetry and slope calibration.
Automatic Buffer Recognition
..............pH4.00, pH6.88, pH7.00, pH9.23 & pH10.06.
Any other can be entered during calibration.
Sensor Asymmetry Range
....................-1.00 to 1.00 pH
Sensor Slope Range
...............................85.0 to 105.0 %