135%).
�
Unstable or
inaccurate
readings.
2. Incorrectly fitted
membrane.
3. Sensor is empty.
4. Sensor is faulty.
flush the filling solution.
Membrane should be smooth
and convex with no wrinkles.
Re-fit membrane if
necessary.
Replace membrane and re-
fill sensor.
Return sensor to factory for
repair or replacement
Blackened Silver
anode
Sensor has been exposed to
sulphides or other chemical
poisoning.
Return to the TPS factory for
cleaning and service.
Tarnished or
scratched Gold
cathode.
Sensor has been chemically
poisoned or physically
damaged.
Return to the TPS factory for
cleaning and service.
Meter reads
OVR
ppM
or
OVR%
.
1. Sensor has not yet
polarised.
1. Sensor is faulty
Wait for 2-3 minutes for the
sensor to polarise after the
Aqua-DY
is switched on.
Return sensor to factory for
repair or replacement.
For complete details on the operation, care and maintenance of your
YSI dissolved oxygen sensor, see the separate leaflet that is
supplied with the sensor.
Page 21