background image

TPP Pipettor Turbo-Fix                              Operating instructions

155923_V07

Summary of Contents for Pipettor Turbo-Fix

Page 1: ...TPP Pipettor Turbo Fix Operating instructions 155923_V07 ...

Page 2: ...nts normatifs ci après cumple las documentos normativos soddisfa le normative seguenti Standards for Canada and USA Trasadingen October 25 2013 TPP Pipettor Turbo Fix 2006 95 EC Low voltage equipment 2004 108 EC Electromagnetic compatibility 2011 65 EC Restriction of Hazardous Substances 2002 96 EC Waste Electrical and Electronic Equipment 2009 129 EC Eco Design Directive 2006 66 EC Battery Direct...

Page 3: ...Operation 7 5 Maintenance 9 6 Technical Data 10 Imprint 2013 TPP Techno Plastic Products AG Supplier Customer service Please contact your local TPP Techno Plastic Products AG representative see www tpp ch or contact info tpp ch TPP Techno Plastic Products AG CH 8219 Trasadingen Switzerland T 41 52 687 01 87 F 41 52 687 01 77 info tpp ch www tpp ch ...

Page 4: ...lines such as the use of protective clothing and safety goggles Never point a pipette in anyone s direction 3 Avoid pipetting of liquids whose vapours could attack the materials PA polyamide POM polyoxymethylene FPM fluor rubber NBR nitrile rubber CR chloroprene silicone Corrosive vapours could also damage metallic parts inside the device 4 Use an original TPP mains adapter only and protect it fro...

Page 5: ...all mount Sterile filter spare Quick Start Guide 2 2 Overview of TPP Pipettor Turbo Fix 1 Aspiration button 2 Dispensing button 3 Thumb wheel to set maximum speed 4 Battery charge indicator 5 Socket for mains adapter cable 6 Handle 7 Nose piece 7 Nose piece 7a O ring 7b Top half of nose piece housing 7c Filter rubber 7d Sterile filter 7e Pipette mount 7f Bottom half of nose piece housing 1 2 4 5 7...

Page 6: ...ounting of the wall mount The wall mount serves to park TPP Pipettor Turbo Fix To mount the wall mount remove the protective foil from the adhesive tape at the back of the holder Hold it with the sign up facing upwards and press it to the desired place Make sure that the surface onto which the wall mount is mounted is smooth clean and grease free Wait 24 hours before using the wall mount for the f...

Page 7: ... faster for large pipettes To empty the pipette by gravity force press the dispensing button only slightly in order to avoid reaching the trigger point where the pump starts running Gravity dispensing is used for to deliver TD pipettes that are not of the blow out type blow out pipettes have two thin rings or a frosted band around the neck TPP Pipettor Turbo Fix is featured with a TURBO mode Plug ...

Page 8: ...dirty Replace filter The nose piece 7 is not tight Tighten the nose piece or replace defective parts The battery is discharged battery charge indicator flashes red Charge the battery The battery is missing Insert the battery or connect the instrument to the mains adapter The battery is defective Replace the battery The battery is wrongly inserted Insert correctly note polarity and Reduced operatin...

Page 9: ...ds TPP Pipettor Turbo Fix After maintenance work perform a leak test to ascertain correct functioning of TPP Pipettor Turbo Fix liquid should not leak out of a filled pipette before the dispensing button is pressed 5 2 Equipment disposal TPP Pipettor Turbo Fix device must not be disposed of with unsorted municipal waste Dispose of TPP Pipettor Turbo Fix in accordance with the laws and regulations ...

Page 10: ...h Typical charging time 3 5 hours Charging cycles 500 1000 when charging as indicated Running time at least 5500 cycles of aspiration and dispensing of 25 ml Mains adapter Input 100 240 VAC 50 60 Hz Output 17 VDC 10 180 mA 10 Materials Housing PA Nose piece housing POM Pipette mount Silicone Filter rubber Silicone Dimensions H x W x D 125 x 130 x 35 mm Weight 195 g Ambient conditions Operation 5 4...

Reviews: