CALIBRATION
Turn on the PGM100 until the working display is illuminated. Wait for 2 minutes prior to
beginning the calibration process.
1. Prepare the approved gas cylinder and regulator with the calibration adapter attached.
The flow should be 200-400 cc/min.
2. Push and hold the Zero button until the display reads "GAS".
3. Attach the PGM100 calibration adapter.
4. Start the gas flow.
5. Push and release the Zero button. A changing number will flash as will the CAL icon.
6. Successful calibration is indicated when only the gas concentration is displayed. The
CAL icon will no longer be illuminated.
7. Bad calibration will be indicated by the word "BAD" and the CAL icon on the display.
Press the power button to acknowledge. (Retry calibration)
8. Successful calibration will be logged as an Alarm event in the log.
BUMP/CALIBRATION using SCal-100
(Sold Separately)
1. Prepare SCal-100 Station.
2. Turn PGM100 On and allow to Zero.
3. Install PGM100 into SCal-100 Station.
4. Remove PGM100 when test is complete