RECALLING DATA
The recall feature allows you to display stored data or retrieve the last date of
calibration.
Recalling Saved Data or Calibration Date
1. Press the Recall button once.
2. The display will show “Stor” and “Cal”. Use the Arrow buttons
to select the desired function.
To cancel recalling information press
the Recall Key and use the Up/Down Arrow Keys to select “Yes” and
press the Scroll/Enter Key.
Stor -
Selecting “Stor” allows you to display the information in a
memory location.
1. Press the Recall button once. The display will show “Stor” and
“Cal”. Using the Arrow buttons select “Stor” by making it blink.
Press the Scroll/Enter button.
2. “Addr” will be displayed and a number from 0 to 49 will
flash.
3. Using the up and down arrows, select the memory location of
the readings you want to display.
4. Press the Scroll/Enter key once. The 708 will display the date
and time screen showing when that reading was stored.
5. Use the up and down arrows to scroll through the data that
was saved.
6. Press the Scroll/Enter button and “End” will be displayed and
“Yes” will flash. Press the Scroll/Enter button if you are finished
recalling data or use the Arrow buttons to select No to recall
data from other locations.
CAL -
Selecting “CAL” allows you to display the last date of cali-
bration.
1. Press the Recall button once. The display will show “Stor” and
“Cal”. Using the Arrow buttons select “Cal” by making it blink.
Press the Scroll/Enter button.
2. “CAL” will blink and the last calibration date will be displayed in
DD/MM.YYYY format.
3. Press the Scroll/Enter button, “End” will be display and YES will
blink. Press the Scroll/Enter button to return to normal
operation.
23
STORING DATA
During testing data can be stored for later retrieval.
When data is saved, all data previously in the address will be overwrit-
ten.When saving pressure/draft readings, select an address different from
the one used to save combustion readings.
1. Press the Store Key.
To cancel storing information press the Store Key
and use the Up/Down Arrow Keys to select “Yes” and press the
Scroll/Enter Key.
2. “Addr” and an address location from 0 to 49 will flash on the display. Use
the Up/Down Arrow Keys to select the desired location to store the data
and press the Scroll/Enter Key.
To cancel storing information press the
Store Key and use the Up/Down Arrow Keys to select “Yes” and press the
Scroll/Enter Key.
3. The data is stored and “End” will display. Use the Up/Down Arrow Keys to
select “Yes” to stop storing data and return to normal operation or select
“No” to store data again. Press the Scroll/Enter Key.
You have just successfully stored a set of readings which can be reviewed
on screen (see page 23) or sent to the IR printer (see page 24).