background image

E. MEASUREMENT TECHNIQUES

   1. Turn the 608(BT) on by pressing the ON/OFF key. The display will initially read “608” along

with “APO” or “- - - -”, and the selected unit of measure. After this initial display the 608(BT*)

is ready for use. 

       *Please see pages 4 to 6 to connect Bluetooth version to the FREE TPI View app. for

enhanced features.

APO = Auto Power Off enabled - the 608 will turn off after 20 minutes

       - - - - = Auto Power Off disabled - the 608 will stay on until turned off

       

       Follow these steps to change the unit of measure and to enable or disable auto power off as

required.

       

       a) With the 608 turned off, press and hold the ON/OFF key until the display flashes “APO”

or “----” and the unit of measure annunciator.

       b) Release the ON/OFF key and then press it repeatedly until the desired unit of measure is

displayed. The unit of measure aanunciators will cycle through mmH2O, mbar, KPa, PSI,

and inH2O each time the ON/OFF key is pressed.

c) Press the ZERO key to enable or disable auto power off. “APO” displayed indicates auto

power off is enabled, “- - - -” displayed indicates auto power off is disabled. Pressing the

ZERO key cycles between these two options.

       d) Once the desired unit of measure and auto power off mode is selected, allow the 608 to

turn off and then turn it back on and proceed with testing. The 608 will turn on in the last

selected unit of measure and auto power off mode until changed.

   

   2. Zero the displayed reading by pressing the ZERO button.

   3. Connect the 608(BT) to the device under test.

   

Additional Features

   

Backlight -

The backlight can be activated to make the display easier to see in 

   

dark environments. Press the BL button to turn the backlight on. The 

   

display will light up and “BL” will be displayed.To preserve battery 

   

life, the backlight will remain on for 15 seconds and then turn off.

   

Case Magnets -

The 608 back case is equipped with magnets and a hook.

   

and Hook

This allows the 608 to be secured during tests.

F.ACCESSORIES

Item #             Description
A009A

............9 Volt Alkaline Battery (included)

A616

..............Soft Carrying Pouch (optional)

A608BT

..........Splashproof Protective Cover (optional)

2

Summary of Contents for 608

Page 1: ...Digital Manometer 608 608BT TestProducts International Inc The Value Leader Visit www tpieurope com ...

Page 2: ...e capability of the 608 BT before connection 3 Check all connections making sure they are secure with no leaks 4 Observe all manufacturers recommendations C General Specifications Sensor Type Temperature compensated solid state Connector Type Standard 1 4 Barb Fitting Range 150 mBar to 150 mBar Units of Measure PSI mBar KPa inH2O Over Pressure 300 mBar Operating Temperature 10 to 50 C 14 to 122 F ...

Page 3: ...y to enable or disable auto power off APO displayed indicates auto power off is enabled displayed indicates auto power off is disabled Pressing the ZERO key cycles between these two options d Once the desired unit of measure and auto power off mode is selected allow the 608 to turn off and then turn it back on and proceed with testing The 608 will turn on in the last selected unit of measure and a...

Page 4: ...ng dam ages to property are not covered Transportation expenses to the company are not covered Some states do not allow the exclusion or limitation of incidental or consequential damages so the above limitations or exclusions may not apply to you To Obtain Warranty Performance Include with the product your name address phone number written description of the problem and proof of purchase date Care...

Page 5: ...ressure Smart Probe It connects to your mobile device and uses the TPI View App to display pressure readings perform Let By Tightness Test Data can be displayed saved logged allocated to a specific job e mailed from within the TPI View app once the 608BT is connected to it 3 Download the View app from either the Apple App Store or the Google Play Store 4 TPI View App As simple as 1 2 3 1 Connect 2...

Page 6: ...rt a different reference name Job Management allows you to create multiple JOBS and attach multiple REPORTS to each JOB Save to an existing JOB or create a NEW one 5 Perform various types of pressure tests such as Live Pressure Fan Pressure Working Pressure even FULL UK Let ByTightness Tests using the FREE TPI View app ...

Page 7: ...6 The 608BT has been tested with the following standards and is in compliance with the council RED directive 2014 53 EU ...

Page 8: ... your account and register your product Benefits include Everything in one place Create service calibration and warranty repair requests Download certificates Contact sales support and customer service Read correspondence from TPI including Service updates Manage your order history Shop Online for accompanying products and spares delivered direct to your door Please do not hesitate to contact us b...