TPC PC2635 Operational Manual Download Page 8

   

5.4   

Draining   of   air   filter:   

To   drain   the   air   filter,   located   in   the   middle   of   the   clear   canister;   first   open   the   

drain   valve,   to   relive   all   air   pressure   in   the   tank.      When   the   air   pressure   reaches   0   

bar,   the   water   in   the   relief   valve   will   automatically   be   released.   

   

5.5   

Air   inlet   filter   core   replacement   

An   air   filter   is   installed   on   the   air   inlet   of   the   motor   to   prevent   dust   from   

entering   the   air   compressor;   it   also   helps   to   reduce   noise.      The   filter   core   will   

need   to   be   checked   and   replaced   regularly.      To   replace   the   air   filter;   open   the   lid   

to   the   lid   on   the   air   filter,   remove   the   old   filter   core,   install   a   new   filter   and   close   

the   lid.   

   

5.6   

Handpiece   Maintenance   

Routine   handpiece   maintenance   including;   oil   and   lubricant   should   prolong   the   

life   of   a   handpiece.   

   

5.7

   

Handpiece   Nozzle   Cleaning   

Weekly   handpiece,   air   &   water   nozzle   maintenance   should   prolong   the   life   of   the   

handpiece.   
   

5.8   

Handpiece   Sterilization   

Spray   the   handpiece   with   water   or   alcohol   before   sterilization.      Disinfect   the   

handpiece   by   soaking   and   wiping   the   exterior.      Bag   and   seal   the   handpiece   and   

sterilize   in   autoclave   at   2   bar   pressure   at   134’C   for   4   minutes;   or   1   bar   of   

pressure   at   121’C   for   18   minutes.   

   

Notice:      

After   each   patient,   all   oral   instruments   must   be   sterilized.   

   

5.9

 

   

Three-­‐Way   Syringe   Sterilization   

First   remove   the   Air/Water   syringe   tip,   before   sterilization.      Disinfect   the   

Air/Water   assembly   by   soaking   and   wiping   the   exterior.      No   sterilization   is   

necessary.      Use   a   barrier   such   as   an   Air/Water   syringe   cover   to   avoid   cross   

contamination.   

Summary of Contents for PC2635

Page 1: ...Portable Dental Unit PC2635 Operational Manual Notice Before Operating the unit please read the manual carefully and reference it for future maintenance procedures...

Page 2: ...n Any damages to any of the wiring cables or attachments are to be replaced immediately The unit should be kept clean and dry there should not be any noticeable oil leaks on any surface of the machine...

Page 3: ...w 2 Symbols 3 Operational Overview 4 Main Functions 5 Operation and maintenance 6 Operating Diagram 7 Standard Configuration 8 Technical Specifications 9 Portable Unit Troubleshooting 10 Air Water Cir...

Page 4: ...s clinics and hospitals for the treatment of oral health Equipped with a self contained air compressor air tank air filter and regulator clean water bottle three way syringe high speed hand pieces tub...

Page 5: ...e air tank is collected at the bottom of the tank where it can be discharged The pressure regulator controls and maintains the air pressure within the air tank You can adjust the air pressure by raisi...

Page 6: ...after filling bottle to tighten turn bottle counter clock wise Notice Before adding water please turn off the net water switch 4 9 Air System Prior to packaging the oil free compressor has passed all...

Page 7: ...erve the pressure gauge on a full tank to see if the reading on the gauge drops If there is no detectable air leak open the air outlet valve next to test the pressure switch When the air pressure drop...

Page 8: ...enance Routine handpiece maintenance including oil and lubricant should prolong the life of a handpiece 5 7 Handpiece Nozzle Cleaning Weekly handpiece air water nozzle maintenance should prolong the l...

Page 9: ...Three way syringe 7 Tank 16 High Speed HP tube 23 Handpiece water adjustor 8 Pressed drain valve 17 Power switch 9 Water Bottle 18 Foot pedal 24 Handpiece air adjustor 7 Standard Configuration 4 Hole...

Page 10: ...of air compressor is too high Touch the shell of the air compressor by hand Cool the air compressor use it until it is cooling 2 Compressor cannot be stared The forward valve fails to work Take off t...

Page 11: ...sure rise or use a manostat The water in water tank has been used up Check the water volume of the water tank Replace the tank 7 The handpiece cannot spray water while in use Air water distributing va...

Page 12: ...10 Air Water Circuit Diagram 10 1 Electrical Circuit 10 2 Air and water circuit 11 Air Water Circuit Diagram 1 PC2635 1 pc 2 Product manual 1 pc...