Portable Dental Unit
PC2635
Operational Manual
Notice: Before Operating the unit, please read the manual carefully and reference it for future maintenance procedures.
Page 1: ...Portable Dental Unit PC2635 Operational Manual Notice Before Operating the unit please read the manual carefully and reference it for future maintenance procedures...
Page 2: ...n Any damages to any of the wiring cables or attachments are to be replaced immediately The unit should be kept clean and dry there should not be any noticeable oil leaks on any surface of the machine...
Page 3: ...w 2 Symbols 3 Operational Overview 4 Main Functions 5 Operation and maintenance 6 Operating Diagram 7 Standard Configuration 8 Technical Specifications 9 Portable Unit Troubleshooting 10 Air Water Cir...
Page 4: ...s clinics and hospitals for the treatment of oral health Equipped with a self contained air compressor air tank air filter and regulator clean water bottle three way syringe high speed hand pieces tub...
Page 5: ...e air tank is collected at the bottom of the tank where it can be discharged The pressure regulator controls and maintains the air pressure within the air tank You can adjust the air pressure by raisi...
Page 6: ...after filling bottle to tighten turn bottle counter clock wise Notice Before adding water please turn off the net water switch 4 9 Air System Prior to packaging the oil free compressor has passed all...
Page 7: ...erve the pressure gauge on a full tank to see if the reading on the gauge drops If there is no detectable air leak open the air outlet valve next to test the pressure switch When the air pressure drop...
Page 8: ...enance Routine handpiece maintenance including oil and lubricant should prolong the life of a handpiece 5 7 Handpiece Nozzle Cleaning Weekly handpiece air water nozzle maintenance should prolong the l...
Page 9: ...Three way syringe 7 Tank 16 High Speed HP tube 23 Handpiece water adjustor 8 Pressed drain valve 17 Power switch 9 Water Bottle 18 Foot pedal 24 Handpiece air adjustor 7 Standard Configuration 4 Hole...
Page 10: ...of air compressor is too high Touch the shell of the air compressor by hand Cool the air compressor use it until it is cooling 2 Compressor cannot be stared The forward valve fails to work Take off t...
Page 11: ...sure rise or use a manostat The water in water tank has been used up Check the water volume of the water tank Replace the tank 7 The handpiece cannot spray water while in use Air water distributing va...
Page 12: ...10 Air Water Circuit Diagram 10 1 Electrical Circuit 10 2 Air and water circuit 11 Air Water Circuit Diagram 1 PC2635 1 pc 2 Product manual 1 pc...