- 16 -
TOTOKU ELECTRIC CO., LTD.
MQSA Quality Control Manual
4 DISPLAY CALIBRATION
4.1 Objective
To ensure that an accurate evaluation of the display monitors will be conducted to
perform an accurate calibration for the reading of the mammographic images on
Monochrome Monitors for Mammography
4.2 Frequency
Annually
4.3 Required Equipment
Monochrome Monitors for Mammography, display monitor
Serial cable for communicating between PC and display monitor (for ME511L)
USB cable for communicating between PC and display monitor (for other models)
Calibration software (Medivisor Nx)
Calibration sensor (Chroma5)
4.4 Procedure
1. Select “TOTOKU Medivisor Nx” > “ Medivisor Nx” from the Start menu of
Windows. Attach the calibration sensor properly on the monitor to calibrate and
click “Calibration”.