1-35
TLPB2 Service Manual Rev. 2.0
2
(1) Monitor the waveform at the test point “TP-G”.
(2) Click the [WMB] button and adjust to the specified level.
(3) Check the voltage at the test points “TP-R” and “TP-B”.
(4) Click [Next] button.
[Note]
“G” and “B” levels also change when the [WMB]
adjustment is performed.
Recommendation value : E8 (min:E0 – max:F5)
3
Input Signal : RGB flicker pattern
(1) Select a red single color screen when adjusting R. Click
the [WM6] button, and adjust the flicker level to minimum in
both + and – directions, taking note of the data values. Then
set data to the mid-point between the two previously noted
values.
(2) Adjust other colors to be the same, using [WM5] for green
and [WM4] for blue.
(3) Click the [Upside Down] button to change projection
mode.
(4) Adjust this mode in the same way.
(5) Click [Next] buttons.
Recommendation value : 40 – 50
4
(1) Monitor the waveform at test point “TP-R”.
(2) Click the [W03] button and adjust to the specified level.
(3) Click [Next] button.
[Note]
A signal level rises when [
–
] is pushed.
A signal level falls down when [+] is pushed.
5
(1) Monitor the waveform at test point “TP-G”.
(2) Click the [W04] button and adjust to the specified level.
(3) Click [Next] button.