6F8A0926
5.4.2 Cable Connection
Connect and install the terminal-treated cables to the terminal block.
*Connect the cables to the terminal block securely. A loose connection may cause incorrect
measurement. After connecting a cable, try to pull it to check whether it has been connected securely.
(1)Referring to
combined converter's manuals of "Connections and Grounding"
, connect each cable to
the terminal block.
Tighten the screws of the terminal block tightly to ensure a secure connection
. A
loose connection may cause incorrect measurement.
After connecting a cable, try to pull it to see whether
it has been connected securely.
Figure 5.5 Connecting a Cable to Terminal Block
(2) Attach the terminal box cover . To keep the seal, tighten securely the cover.
Phillips screwdriver
Terminal block
* The appropriate torque for tightening
the terminal board screw is 1.2 N.m.