Example – clear the log file:
send: lc<CR>
returned: ok <CR><LF>
Command:
ls
Function:
Start Logging
Description: When the command ls<CR> is received by the SC25 unit, the unit will begin to log plate
temperature values at the rate defined by the log period setting (ref le, lm, or l5 commands or front panel
menu command). If the command is not received in the proper syntax, e<CR><LF> will be returned.
Example – start logging:
send: ls<CR>
returned: ok <CR><LF>
Command:
lp
Function:
Stop or Pause Logging
Description: When the command lp<CR> is received by the SC25 unit, the unit will stop logging log plate
temperature values. If the command ls is issued, the unit will continued logging from the last log position.
If the command is not received in the proper syntax, e<CR><LF> will be returned.
Example – stop logging (or pause logging)
send: ls<CR>
returned: ok <CR><LF>
Command:
le
Function:
Set Logging Period to 1 second
Description: When the command le<CR> is received by the SC25 unit, the unit will set the logging period
to 1 second which means that a data point will be stored once every second. If the command is not
received in the proper syntax, e<CR><LF> will be returned.
Example – set to log a data point every second
send: le<CR>
returned: ok <CR><LF>
Command:
lm
Function:
Set Logging Period to 1 minute
Description: When the command lm<CR> is received by the SC25 unit, the unit will set the logging period
to 1 minute which means that a data point will be stored once every minute. If the command is not
received in the proper syntax, e<CR><LF> will be returned.
Example – set to log a data point every minute
send: le<CR>
returned: ok <CR><LF>
Command:
l5
Function:
Set Logging Period to 5 minutes
Description: When the command l5<CR> is received by the SC25 unit, the unit will set the logging period
to 5 minutes which means that a data point will be stored once every 5 minutes. If the command is not
received in the proper syntax, e<CR><LF> will be returned.
Example – set to log a data point every 5 minutes