![TORREY PINES SCIENTIFIC EchoTherm IC20 Operating Manual Download Page 5](http://html1.mh-extra.com/html/torrey-pines-scientific/echotherm-ic20/echotherm-ic20_operating-manual_1134651005.webp)
Torrey Pines Scientific, Inc.
5
IV. LABELS
There are various labels on the body of this unit. Listed below are the labels and their
meanings.
This symbol means: ATTENTION. The INSTRUCTION MANUAL IS TO
BE CONSULTED FOR FURTHER INFORMATION
This symbol means: WARNING. HOT SURFACE.
This symbol means: GROUND OR EARTH CONNECTIONS
V. CAUTIONS
CHILLER/HEATER PLATE SURFACE
The IC20/IC30 Series are capable of chilling and heating the plate surface from -10ºC
(-20C on XT versions) to 100ºC (110ºC on XR versions). Temperatures from 44ºC (110ºF)
to 100ºC (212ºF) are hot enough to burn the skin if touched.
Use extreme caution at all
times.
Never leave your unit accessible to others when it is hot. Never touch the plate
surface unless you are sure it is cool.
Note that there is a red LED on the front panel
marked HOT. This will illuminate when the heater surface goes above 50.0°C (122°F).
ELECTRICAL
The cooling/heating module in the IC20/IC30 Series runs using 12 volts DC at 4.2 amps for
the IC20 and 8.4 amps for the IC20XR, IC20XT & all IC30’s. The units are supplied with a
universal power supply that can take inputs from 100 to 260 volts AC, 50/60 Hz. The units
are supplied with an AC input cord for the power supply. Be certain to use a line cord with
the same rating and of the same type as the one supplied by the manufacturer. Use the
normal care and precaution one would use with any electrical appliance.