background image

SL-D

 SERIES

ISSUED............... AUG.  2003

REVISED ............ 

REPAIR MANUAL

PHOTOGRAPHING DEVICE

TOPCON CORPORATION

P-SLD_OPTION-0-0308-14

Summary of Contents for SL-D Series

Page 1: ...SL DSERIES ISSUED AUG 2003 REVISED REPAIR MANUAL PHOTOGRAPHING DEVICE TOPCON CORPORATION P SLD_OPTION 0 0308 14...

Page 2: ...ONE is connected 9 3 SL D7 SR 53 Fuji S1Pro or S2 Pro FD 21 IMAGEnet is connected 10 4 SL D7 TL 54 JVC KY F70BU FD 21 IMAGEnet is connected 11 5 SL D7 DC 1 FD 21 IMAGEnet is connected 12 6 SL D7 DC 1...

Page 3: ...in those areas should take the repair course sponsored by Topcon 1 3 How to Use This Manual and Precautions 1 3 1 Precautions for repair 1 Carry out repair assembly disassembly and adjustment on opti...

Page 4: ...edure Follow the procedures described in the flow chart below YES NO YES NO Repair is necessary Trouble has occurred Understand the trouble and its cause Normal Repair Inspection Makeup Packing End P...

Page 5: ...insertion port Slit lamp Photographic button Xenon cord Power switch Slit lamp connecting connector IMAGEnet connecting connector Illustration of connector insertion port AC inlet Light volume indicat...

Page 6: ...Name Q ty Name Q ty Xenon relay lens 1 Photographic power supply 1 Background lighting BG 3 1 Power cord FD 21 1 Photographic arm cover 1 Cable support 1 Camera attachment and connecting cable when S...

Page 7: ...nd Functions Filament position Camera mount reference surface Xenon position Slit position Film position SR 53 DC 1 SR 52 CCD TL 54 TL 55 6 3 4 4 3 4 6 3 2 6 3 4 7 1 5 Photographic system Photographic...

Page 8: ...observed image and switches it to the camera side Photographic lens Lens for projecting the observed image Focusing of the observed image Reflection mirror Used to reflect the observed image Diffusio...

Page 9: ...R 2 CN1 CN1 CN1 CN1 CN1 CN1 1 2 3 4 5 6 1 2 3 4 5 6 EHR 10 EHR 10 DE 9S UL IMAGE NET RELEASE R L DE 9S UL B10B EH MU03 4201 8 SW3 S22B1109D CN3 CN3 CN3 CN3 CN3 CN3 CN3 CN3 CN3 CN3 CN3 CN3 1 2 3 4 5 6...

Page 10: ...J13 P13 P13 P13 P13 P13 P13 P13 P13 P13 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 P17 J17 P17 J17 P17 J17 P17 J17 P17 J17 SYNC GND REL GND 12V SOLENOID 1 2 3 4 5 1 2 3 4 5 J18 P18 NIKON D1 CORD 4469186200 J...

Page 11: ...4 5 6 7 8 J13 J13 J13 J13 J13 J13 J13 J13 J13 P13 P13 P13 P13 P13 P13 P13 P13 P13 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 P17 J17 P17 J17 P17 J17 P17 J17 P17 J17 1 2 3 4 5 1 2 3 4 5 J18 P18 BF1011 BF1011...

Page 12: ...N5 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 J13 J13 J13 J13 J13 J13 J13 J13 J13 P13 P13 P13 P13 P13 P13 P13 P13 P13 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 P17 P17 1 2 P15 P15 P15 P15 P15 P15 P15 P15 P15 36 10 2 3...

Page 13: ...CN2 CN5 CN2 CN5 CN2 CN5 CN2 CN5 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 J13 J13 J13 J13 J13 J13 J13 J13 J13 P13 P13 P13 P13 P13 P13 P13 P13 P13 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 P15 P15 P15 P15 P15 P15 P15...

Page 14: ...4479388000 FD 21 Xe power supply P10 P10 P10 P10 51021 0400 P11 P11 P11 P11 J10 J6 J10 J10 J10 USB CN 51021 0400 IF CN XE_FLASH EXT_TRIG EYE_L R GND POWER D D GND CVBS GND J11 J11 J11 J11 1 2 3 4 1 2...

Page 15: ...J11 J11 J11 J11 1 2 3 4 1 2 3 4 1 J4 J18 J19 J5 J8 J13 J7 J4 P4 12V 1A 1 1 J4 P4 2 2 2 3 4 1 2 3 4 J12 1 J12 2 51021 0200 VIDEO OUT CN P12 1 P12 2 J S SW 2 1 3 CN7 P11 CN7 P11 CN7 P11 CN7 P11 CN7 P11...

Page 16: ...Focus position ing tool SR 53 1 Shift of screen center Project the 0 2 aperture on the test bar and measure with the focus positioning tool 1mm in up down right left Positioning of optical axis for SR...

Page 17: ...drum to 16 and project the 10 aper ture on the test bar Remove the aperture frame 45520 1112 Attach the focus positioning tool to the camera mount unit Lower the return mirror with your finger and mak...

Page 18: ...h the rubber cap Project the 0 2 aperture on the test bar and as looking into the focus positioning tool check the amount of shift When the optical axis is shifted carry out posi tioning of optical ax...

Page 19: ...ct the 0 2 aper ture on the test bar Remove the aperture frame 45520 1112 Attach the focus positioning tool to the camera mount unit As looking into the focus positioning tool check the amount of shif...

Page 20: ...the 3 screws CR2 4S securely and apply the screw lock When the optical axis is shifted by 1mm or more in the up and down direction Loosen the 1 piece 1N2 6 adjust 1 piece U2 6 6S and bring the optical...

Page 21: ...lustration Procedure After adjusting accuracy tighten 1N2 6 securely and apply the screw lock Adjust 1 piece U2 6 6S and bring the optical axis to the center 0 2 aperture Scale line 8 P SLD_OPTION 0 0...

Page 22: ...6 and project the 10 aper ture on the test bar Attach the focus positioning tool to the camera mount unit Make sure that the scale line on the focus posi tioning tool is fit to the focus on the test b...

Page 23: ...h the rubber cap Project the 0 2 aperture on the test bar and as looking into the focus positioning tool check the amount of shift When the optical axis is shifted carry out posi tioning of optical ax...

Page 24: ...project the 0 2 aper ture on the test bar Remove the aperture frame 45520 1112 Attach the focus positioning tool to the camera mount unit As looking into the focus positioning tool check the amount o...

Page 25: ...ed by 1mm or more in the right and left direction Make adjustment with the 3 push screws 6SU3 6S and the 3 pull screws CR3 12S After adjusting accuracy apply the screw lock to the 6 screws 6SU3 6S CR3...

Page 26: ...of medical equipment Only qualified persons should carry out the following repairs 2 Special tools are necessary for repair Focus positioning tool When this tool is necessary for repair owing to unavo...

Page 27: ...1004 2 Removal of the quick return mirror holder unit Remove the cover 1004 Remove the E ring Remove the connecting plate 1017 from the shaft Remove the 4 screws CR2 3S Remove the quick return mirror...

Page 28: ...manual for TL 54 55 Replace the unit for DC 1 2 Photographic focus is not clear Eyepiece visual acuity is not cor rect Attach the test bar and adjust the eye piece visual acu ity P 22 23 Focus is not...

Page 29: ...k mirror operate normally The Xenon lamp does not light sometimes NO YES Inspection 2 Does the Xenon lamp light It must be repaired in factory NO YES When the light volume indicator is changed does th...

Page 30: ...28 Inspection 1 Checking order Contents of check Has the fuse blown out Replace the fuse 1 YES NO The power switch does not light It must be repaired in factory P SLD_OPTION 0 0308 14...

Page 31: ...lamp is faulty Replace it YES The Xenon lamp does not light sometimes NO Check the wiring 4 Does the Xenon lamp light Is the photographic unit normal Check the wiring 4 NO YES NO Does it light YES It...

Page 32: ...2 Checking order Contents of check Is the joystick switch normal The joystick switch is faulty Replace it YES Check the wiring 9 NO YES It must be repaired in factory NO Does it light Normal P SLD_OP...

Page 33: ...check Is the joystick switch normal The quick mirror does not operate normally NO YES Is the quick mirror normal NO YES The joystick switch is faulty Replace it Check the wiring 9 Normal SR 52 must be...

Page 34: ...lty Replace it YES The commercial camera does not operate normally NO Check the connection Refer to the assembly instruction manual Does the commercial camera operate normally NO YES Is the joystick s...

Page 35: ...he way such a part is shown in Via Check it at the same time Refer to Electric System wiring Diagram No Applicable matter Checking interval From Via To Remarks 4 Disconnection Connector P13 Connector...

Page 36: ...34 2 3 5 Electric parts arrangement 1 FD 21 2 SR 52 3 Xenon lamp Fuse 1 SOL Connector Connector 2 3 4 5 P SLD_OPTION 0 0308 14...

Page 37: ...ol Order No Rating Remarks Fuse F1 F2 44790 5501 5TT10 Bell Fuse Quick mirror solenoid SOL1 44790 1023 Connector J17 44642 31520 HR 12 10R 5SD Hirose Connector J18 44691 15050 R03 RB5F Tajimi Xenon la...

Page 38: ...ibrations 0 1mm for SR 52 53 Focusing Positioning of optical axis Diopter tele scope Observation mag nification 3 4 Focusing Tester type that can measure voltage electric current and resistance Radio...

Page 39: ...ch Industrial Complex SINGAPORE 128381 Phone 62780222 Fax 62733540 www topcon com sg TOPCON SOUTH ASIA PTE LTD Excella Business Park Block C Ground 1st Floor Jalan Ampang Putra Taman Ampang Hillir 551...