12
ASSEMBLY PROCEDURE
ASSEMBLY PROCEDURE
ASSEMBLY PROCEDURE
No special tools are required. Included tools are:
• Large screwdriver
• Small screwdriver
• Phillips screwdriver
• Wrench (not supplied with unit model slit lamps)
(1) Mounting the Table
(a) To attach the table on the instrument table AIT-15, use the four 8×24mm bolts with locking
washers.
* Raise the table to allow the bolts to pass through the mounting flange.
* Place the table on the mounting flange of the instrument table and screw the bolts into the
mounting bracket. The controls of the power supply should face the practitioner. Tighten the
bolts securely with the included wrench.
(b) To secure the unit model table to the ophthalmic unit.
* Peel off the tape which secures the plastic washer to the mounting bracket's shaft.
* Insert the shaft of the mounting bracket into the hole of the arm on the ophthalmic unit. The
plastic washer should be between the mounting bracket and the arm.
Note: The power supply is attached to the left side of the unit top (practitioner's view). If the
ophthalmic stand is located to the left of the ophthalmic chair, the power supply must
be relocated to the right side of the unit top to prevent interference with the arm. In this
instance, remove the four wood screws which attach the power supply to the unit top,
re-position the power supply in a similar position to the right side of the unit top, and re-
attach the power supply with the four screws.
Washer