Topcon IS-5000 User Manual Download Page 1

 

 

 

INSTRUMENT STAND 

IS-5000

Summary of Contents for IS-5000

Page 1: ...INSTRUMENT STAND IS 5000 ...

Page 2: ...fe condition if anything is found wrong with the unit 4 After Operation Never hold and pull the connecting cord for disconnecting plugs Excessive force may be applied and cause damage to the inside connections If the unit has not been used for a long period of time check the operation and safety features prior to using it again SAFETY CONCERNS The IS 5000 Stand conforms to UL 60601 1 and CSA C22 2...

Page 3: ...argeable Instrument Handles 6 3 3 4 Chart Projector optional 6 3 4 Power Up 7 4 SYSTEM OPERATION 8 4 1 Main Power Switch 8 4 2 Lamp Switch 8 4 3 Chair Control Switch 8 4 4 Chair Auto Return Button 8 4 5 Auxiliary Receptacle Button 8 4 6 Charging Wells 8 5 ADJUSTMENTS 9 5 1 Refractor and Slit Lamp Arm 9 5 1 1 Counterbalance Adjustment 9 5 1 2 Brake Adjustment 9 5 1 3 Counterbalance Brake Adjustment...

Page 4: ...Instrument Stand Main Components Figure 1 shows the primary components and key features of the IS 5000 Instrument Stand Use this figure as a reference Figure 1 IS 5000 Instrument Stand Pole Base Assembly Refractor Arm Overhead Lamp Slit Lamp Arm Control Panel ...

Page 5: ...strument Wells Battery Voltage 3 5 v Charge Rate Initial 150 mA Trickle 15 mA Electrical Voltage Current Frequency US Versions 120 VAC 10 5 A 60 Hz Non US Versions 240 VAC 10 3 A 50 60 Hz Weight Net weight 288 lbs 131 kg Environmental Operating Conditions Ambient Temperature Range 50 F to 104 F 10 C to 40 C Relative Humidity Range 30 to 75 Atmospheric Pressure Range 10 1 to 15 4 psi 700 to 1060 hP...

Page 6: ... the plastic front cover Use protective padding if needed 3 Unscrew the seven 7 Phillips screws from the shroud cover of the IS 5000 and then carefully remove the cover 4 Back out the two 2 set screws from the black steel tube bracket and remove the plastic tubing out of the bracket Insert pole into the steel tube bracket and slide it down until it stops Rotate the pole until it drops into the key...

Page 7: ...igure 3 2 Instrument arm mounting locations 7 Optional Attach the chart projector arm into position on the pole and secure in place 8 Unpack the overhead lamp from the accessory box Connect the lamp cable connector to its mating connector from the cable push the cable down the pole and then mount the lamp in to the top of the pole Secure the lamp onto the pole by tightening the two 2 screws provid...

Page 8: ...o a grounded wall outlet NOTE Using a non Topcon chair may require an adapter see Table 1 for pin functionality The chair supplies AC power to the IS 5000 and the IS 5000 acts as a switch and returns the power to the chair to make it go up and down Pin Function Pin Function 1 Earth Ground from chair 6 N C 2 AC Neutral from chair 7 Down Limit Switch engaged 3 AC Hot from chair 8 Up pump engaged 4 N...

Page 9: ... S arm by placing the pivot pin on the bottom of the tabletop into the S arm bushing Lock the Slit Lamp on the arm by tightening the knob Plug the cord into the transformer box Refer to section 5 to counterbalance the arm 3 3 3 Rechargeable Instrument Handles The instrument console is supplied with three recharging wells for use with rechargeable instrument handles Simply inserting them into the w...

Page 10: ...ions of all the instruments install the back cover of the IS 5000 To install the removable back shroud cover carefully push the cover forward into place and securely install the seven 7 Phillips screws Peel and discard the protective clear tape from floorplate Well Cable Up Down Switch Low Voltage Lamp Fuse 3A Automotive with LED 120V 240V units O H Lamp Chair Fuse 5A 250V Normal Blow 120V units 3...

Page 11: ...ht can be controlled from the control panel Press the or UP or DOWN button on the control panel to adjust the chair to the desired height 4 4 Chair Auto Return Button Pressing the AUTO button on the control panel will return the chair lift to its lowest position 4 5 Auxiliary Receptacle Button The auxiliary buttons AUX 1 2 control the power to the auxiliary receptacles located to the right of the ...

Page 12: ...p Arm At the instrument end of the arm there is a handle which activates the brake This handle locks the three arm functions simultaneously the instrument end of the arm the pole end of the arm and the counterbalance mechanism Each of these three brakes can be adjusted independently of the others At the instrument and pole ends of the arm are two 2 Hex set screws The small setscrew adjusts the bra...

Page 13: ...rms may tend to drift slightly Increasing the drag on the arm will eliminate a slight drift Drag is adjusted by turning the larger of the two 2 set screws located at both ends of the arm If drifting persists use a level and check that the stand pole is vertical Just a small amount of vertical offset can cause a significant amount of arm drift Figure 5 Counterbalance Adjustments for Refractor Arm ...

Page 14: ...art below a Turn power switch off and unplug the power cord plug from the wall outlet b Push down and turn the fuse holder cap counterclockwise then pull the fuse retainer out c Replace the blown fuse s with a new one refer to Table 2 Push in and turn the fuse holder cap clockwise to replace it Location Fuse 100 120 Volt 220 240 Volt Base Chassis Primary 5A 250 v SB 3A 250 v SB Base Chassis Second...

Page 15: ... questions about this instrument Model Name IS 5000 120V or 240V Serial Number This is provided in the product rating label on the back of the base assembly Period of use Please inform us of its date of purchase Defective condition Please provide us with as much details as possible IS 5000 Topcon Instrument Stand USER MANUAL 2014 version 2031005555 Revision B Date of issue April 14 2014 ...

Page 16: ...2031005555 Revision B 04 14 Topcon Medical Systems Inc 111 Bauer Drive Oakland NJ 07436 800 223 1130 www topconmedical com ...