114
GENERAL INFORMATION ON USAGE AND MAINTENANCE
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC standard (IEC 60601-1-2 Ed.3: 2007).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement
parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment. If adja-
cent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify normal
operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than those
specified may result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYS-
TEM.
Item
Article code
Cable Shielded
Ferrite Core
Length (m)
AC power cable
41840 50100
No
No
1.5
LAN cable
47005 35100
No
No
2.0
RGB cable
-
Yes
Yes
1.9
USB cable (for USB keyboard)
-
Yes
Yes
2.1
USB cable (for mouse)
-
Yes
No
1.8
USB cable 1
40422 74100
Yes
No
2.0
USB cable 2
40422 74100
Yes
No
2.0
Stereo jack cable
-
Yes
No
1.5
DSC power cable
47003 75300
No
Yes
1.3
DSC USB terminal cable
47003 71700
Yes
No
1.5
Guidance and manufacturer's declaration - electromagnetic emissions
The DRI OCT-1 Model Triton is intended for use in the electromagnetic environment specified below.
The customer or the user of the DRI OCT-1 Model Triton should assure that it is used in such an envi-
ronment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The DRI OCT-1 Model Triton uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions
CISPR 11
Class B
The DRI OCT-1 Model Triton is suitable for use in all estab-
lishments including domestic and those directly connected
to the public low-voltage power supply network that sup-
plies buildings used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies