53
3.7 Area Correction Factor
The number of the currently set area correction factor group is displayed.
Display Method
☞
「
3.1.2 Data/Setting Item Display
」
The area correction factor corrects measurement values in an arbitrarily specified area of a xy
chromaticity diagram or u'v' chromaticity diagram.
・
To register an area correction factor, it is necessary to use software CS-900A, or create
communication software that references a command list.
Command List
☞
「
4. Communication with a PC
」
Area correction factors can be registered for a total of 10 groups to the BM-7AC. Groups 1 to
5 are assigned to area correction factors for the xy chromaticity diagram, and groups 6 to 10 are
assigned to area corrections factors for the u'v' chromaticity diagram. Also, it is possible to set
5 chromaticity areas in 1 group. In the chromaticity area setting, the correction factors are
specified by KX, KY and KZ, and the minimum and maximum areas of application of correction are
specified by x, y, or u' ,v', and minimum Luminance are specified by L.
*
FACTOR
GROUP
*
OFF
Memo
Specified color area
Area correction group
(measurement device side)
G1
Color area 1
G2
Color area 2
Color area 3
Color area 4
G5
Color area 5
G6
G7
G10
x,y
u',v'