7) Do the step (4) and make the image clear which is 

observed through the left eyepiece by turning the left 

diopter adjusting ring 

. (Fig.3)

4-5

 

Adjust the interpupillary distance

1)

 

Adjust the prism housing along the direction 

of arrowhead of the 

 Fig.4 till the observation is comfortable.

4-6

 

Use Eyepiece shields

                           

1) For user who does not wear glasses, hold the diopter-

adjusting ring to prevent them from rotating and turn the 
eyepiece till the eyepiece shields fit the observer well.

 

                            

2) For user who wears glasses, take the eyepiece shields 

off before

                           observation

4-7

 Mount and Remove the Optional Eyepiece Micrometer

1) Turn and remove the mounting ring

 from the 

eyepiece.(Fig.5)

2) Clean the eyepiece micrometer

 

and mount it to the 

mounting  

 ring with the inscription side downward.
3) Gently twist the mounting ring with the eyepiece micrometer into

the eyepiece till tightening

 

securely.

                  4) To remove the eyepiece micrometer, take down the mounting 

ring

 

by twisting and take out of the micrometer, and 

then wrap it.

 

 

4-8

 

Install the illumination device

1)Insert the illumination device

 in the bracket with the 

protrudent

 

Fig.4

Fig.5

Summary of Contents for Tekscope SZM

Page 1: ...Contents 1 Before use 1 2 Nomenclature 2 3 Assemblage 5 4 Operation 7 5 Configuration 13 6 Technical parameter 15 7 Troubleshooting 17 1...

Page 2: ...of the big camera 1 2 MAINTENANCE 1 All lenses must be kept clean Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens tissue Fingerprints or...

Page 3: ...e 2 1 SZM 45B1BL2 Eyepiece shield Eyepiece Diopter adjusting ring Head Illumination device Zoom control knob Focusing Lamp lock screw Glass stage Clip Base Top variable light control Fuse Bottom varia...

Page 4: ...2 2 SZM T1BL2 Diopter adjusting ring Eyepiece shield Zoom control knob Eyepiece Head lock screw Head Focusing knob Glass stage Base Clip...

Page 5: ...3 A s s e m b l a g e 3 1 SZM 45B1BL2 Eyepiece shield Eyepiece Head Head lock screw Illumination device Clips Glass stage Base...

Page 6: ...3 2 SZM 45T1BL2 SZM PH OR SZM CTV...

Page 7: ...4 O p e r a t i o n 4 1 Use the glass stage PK mount Photo eyepiece Eyepiece shield Eyepiec e Head C mount Illumination device Clip Glass stage Base...

Page 8: ...focusing bracket from slipping down by its weight during the observation Fig 2 4 3 Set the specimen slide 1 Set the specimen on the center of stage plate If necessary clamp the slide with the clips 2...

Page 9: ...the observer well 2 For user who wears glasses take the eyepiece shields off before observation 4 7 Mount and Remove the Optional Eyepiece Micrometer 1 Turn and remove the mounting ring from the eyep...

Page 10: ...photo eyepieces socket of the tri ocular 2 Connect the PK mount adapter with the photo eyepiece and then tighten the lock screw Fig 8 Fig 8 Fig 9 Fig 10 4 11 Adjust the CTV 1 Adjust the CTV to a suit...

Page 11: ...D board If they have been installed this step can be omitted 3 Double click the icon of the software and then the video window will appear You can set the size of the window according to your linking...

Page 12: ...function automatically analyzing function such as slightness body grain body line body and so on The outcome can be kept in data and can be made into chart and so forth 4 19 Use the white balance 1 T...

Page 13: ...into the jack thoroughly Fig 14 4 Recover the stage plate Fig 15 Note Before replacing the lamps turn off the power first Avoid violence while the lamp is plugged into the jack Fig 15 4 22 Replace the...

Page 14: ...MEWh15X15 Eyepieces SZMEHWh20X10 Binocular SZM7045 O O O O O Tri ocular SZM7045TR SZMAO0 5 165mm O O O O O SZMAO1 5 45mm Auxiliary objective SZMAO2 30mm SZM A1 O O O O Focusing arm SZM A3 O SZM B6 O S...

Page 15: ...X10 Binocular SZM7045 Tri ocular SZM7045TR O O O O O SZMAO0 5 165mm O O O O O SZMAO1 5 45mm Auxiliary objective SZMAO2 30mm SZM A1 O O O O Focusing arm SZM A3 O SZM B6 O SZM BL2 O SZM STL1 O SZM STL2...

Page 16: ...1 10 5X 19 14X 14 3 10X 20 45X 4 4 22 5X 8 9 67 5X 3 90X 2 2 10 5X 21 4 5 25X 42 8 15 75X 14 3 21X 10 7 15X 15 67 5X 3 3 33 75X 6 7 101 25X 2 2 135X 1 7 14X 14 3 7X 28 6 21X 9 5 28X 7 1 20X 10 90X 2...

Page 17: ...irt appears in the field of view Dirt on the surfaces of eyepiece Clean the surface 3 Image is not clear Dirt on the surfaces of the objectives Clean the objectives Diopter adjustment is not correct R...

Page 18: ...b will burn out soon Replace with a new one 13 The bulb flickers or the brightness is unstable The bulb was not inserted correctly Insert it correctly 7 2 Video troubleshooting Trouble Cause Remedy 1...

Page 19: ...he window of the view does not come out The input signal does not accord with the signal which is chosen in the driver of the A D board Choose the correct signal model which match the input signal 7 T...