  Operation manual of TH2685/TH2686 Electrolytic Capacitor Leakage Current Meter 

3. Instruction of panel function 

Label  

Name of key 

Function  

Brand and model   

 

Leakage current 

display window   

Display the value of measured leakage current adopting 3 
1/2 digit display, the unit includes    nA

μ

A

mA 

Test voltage display 

window  

Display the value of measured voltage adopting 3digit 
display  

Sorting indication   

When the leakage current value of tested capacitor is 
smaller than the set one.

PASS indicator lights up

where 

as, FAIL indicator lights up with alarm indication.   

Leakage current 

setting switch   

The value is the qualified-setting leakage current value. 
When the tested leakage current value is smaller than the 
set one, it means the product is qualified, whereas is 
unqualified. There are 3 digits of diver plate value with 
format of AB

×

10

n

nA

if the setting is 324

the set leakage 

current value is 32

×

10

4

nA=320

μ

Charge time setting 

switch  

There are 3 digits of diver plate value with format of ABC 

×

 0.1 s

if the setting is 324

then the charge time is 324 

×

 

0.1=32.4s

the charge of  tested capacitor is performed 

automatically according to the setting of drive plate

when 

the charge finishes, the instrument turns to the 
measurement status automatically.   

Voltage 

transformation key 

TH2685 can be selected to 75V or 200V bin freely

 

and TH2686 can be selected to    200V or 5 00V bin 

Correction key 

Zeroing status

Zeroing indicator lights and the instrument 

can be performed to open correction. After pressing start, 
the instrument performs 5 bin open correction, when the 
correction finishes press correction key, and the instrument 
returns to the measurement status.   

Lock key   

Locking

Locking indicator lights now and the display data 

is the Max. value of capacitor leakage current when the 
charge finishes.   

Unlocking

The display data is the actual tested value of 

capacitor leakage current under the current status.   

10 

Discharge key 

On the status of charge

Press discharge key to perform 

discharge on the tested capacitor. 
On the status of measurement

Press discharge key to 

perform discharge on the tested capacitor. 
On the status of discharge: Press discharge key to perform 
discharge on the tested capacitor. 
 

11 

Start key 

On the status of correction

start key to perform open 

correction on the instrument. 
On the status of discharge, start key to perform 