58
DEFault SLOW
SLOW Set the test speed as SLOW
FAST Set the test speed as FAST
PLAC4 Set the display digits as 4 1/2
PLAC5 Set the display digits as 5 1/2
Query:
:NPLCycles?
Query the current test speed
Description:
The command is used for the basic measurement functions (except Frequency and Period) and test
speed (display digits).
:RANGe Commands
:[UPPer] <n>
Command syntax:
: CURRent:AC:RANGe[:UPPer] <n>
Set measurement range for ACI
: CURRent:DC:RANGe[:UPPer] <n>
Set measurement range for DCI
Command Parameter:
<n> = (
1m,10m,100m01,10)
A
ACI
(1m,10m,100m01,10)
A
DCI
DEFault 1mA (ACI, DCI)
MINimum
1mA (All functions)
MAXimum 10A
Query:
:RANGe[:UPPer]? Query
measurement range of the current functon.
Description:
This command is used to manually select the measurement range for the specified measurement
function.
:AUTO <b>
Command syntax:
: CURRent:AC:RANGe:AUTO <b>
Set auto range for ACI
: CURRent:DC:RANGe:AUTO <b>
Set auto range for DCI
Command parameter:
<b> = 1 or ON
Enable auto range
0
or
OFF Disable
auto
range
Query:
:AUTO?
Query
auto
range
(ON
or
OFF)