57
6.3.5 RESIister subsystem
The commands in this subsystem are used to configure and control resistance measurement function
and are summarized in Table 6-5.
Table 6-5
RESIister Subsystem Commands Summary
Command
Function Description
Default
:RESistance
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure resistance
Set integration rate (line cycles: 0.5 to 2)
Query line cycle integration rate
Path to configure measurement range
Select range (0 to20e6)
Query range
Enable or disable auto range
Query auto range
Specify reference (0 to 20e6)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
1
20e6
ON
0
OFF
Speed Commands
: NPLCycles <n>
Command Systac:
:RESistance:NPLCycles <n>
Set NPLC for
Ω
Command Parameter:
<n> = 0.5 to 2
Set power line cycles per integration
DEFault
1
MINimum
0.1
MAXimum 10
Query:
:NPLCycles?
Query programmed NPLC value
Description:
The integration period (measurement speed) for the basic measurement functions (except
Frequency and Period) is set using the:NPLCycles command. NPLC(Number of Power Line Cycles)
expresses the integration period by basing it on the power line frequency. For example, for a PLC of
1, the integration period in seconds would be1/60 (for 60Hz line power) which is 16.67 msec.