89
sequence in the programmable mode. Returns <value> with a value equal to
0~999.
6.3.5.34
:FUNC: SS
Command Syntax: :FUNCtion:SingleStep ON or :FUNCtion:SingleStep 1
Shorthand for command: :FUNC:SS ON or :FUNC:SS 1
Command description: This command is used to turn on the single step test function in the programmable
mode. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SingleStep OFF or :FUNCtion:SingleStep 0
Shorthand for command: :FUNC:SS OFF or :FUNC:SS 0
Command description: This command is used to turn off the single step test function in the programmable
mode. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SingleStep?
Shorthand for command: :FUNC:SS?
Command description: This command is used to query the single-step test status in the programmable
mode. Returns 1 when the single step test function is turned on and 0 when the
single step test function is turned off.
6.3.5.35
:FUNC: VOLT:LMT
Command Syntax: :FUNCtion:VOLTage:LilMT <value>
Shorthand for command: :FUNC:VOLT:LMT <value>
Command description: This command is used to set the output voltage limit protection in manual mode.
<value> indicates the voltage limit value, ranging from 5.0 to 50.0V. This value
represents the allowable difference between the set voltage and the output
voltage.
Command Syntax: :FUNCtion:VOLTage:LilMT?
Shorthand for command: :FUNC:VOLT:LMT ?
Command description: This command is used to query the voltage output limit value in manual mode, and
returns <value>, which is equal to 5.0~50.0.
6.3.6
FETCH Subsystem Commands
The FETCH subsystem command is mainly used to query the values of output parameters such as voltage,
current, power, ampere peak, power factor, and crest factor in manual mode. The FETCH subsystem command
structure is shown in Figure 6-8.