3-5
■
3.1.5 Positioning by Touch Alignment
When moving an element up, down, left and right using Touch
Alignment, touch the panel and release your finger immediately. Do
not press the panel continuously.
Use Touch Alignment or the joystick to position the patient's eye on the screen.
Lightly touch the pupil center of the patient’s eye displayed on the screen. The
measuring head moves so that the patient's eye is positioned in the center of the screen.
When the center of the cornea enters the alignment ring with Auto Alignment turned on,
focusing in the X and Y axes automatically starts.
Perform alignment in the Z axis. While pressing the center of the screen, the
measuring head advances towards the patient. Touch the measuring head retreat
button to retract the measuring head towards the physician.
When the focus indicator appears while Auto Alignment is activated, focusing in the
Z axis starts automatically. When alignment conditions become optimal, the
alignment OK mark is displayed.
Summary of Contents for MR-6000
Page 2: ......
Page 24: ... 1 18 This page is intentionally left blank ...
Page 30: ... 2 6 This page is intentionally left blank ...
Page 46: ... 4 2 This page is intentionally left blank ...
Page 56: ... 6 4 This page is intentionally left blank ...
Page 58: ... 7 2 This page is intentionally left blank ...
Page 67: ...8 9 This page is intentionally left blank ...
Page 68: ......
Page 70: ...0120 ...