TOELLNER TOE 9260 Series User Manual Download Page 3

 

 

 

TOELLNER Electronic Instrumente GmbH 

1. 

Introduction 

The Electronic Switch TOE 9260 is a device for time-based turning on and off of load and 
signal currents. This device provides an electronic, unidirectional power switch as well as 
four electronic bidirectional signal line switches. The power switch incorporates a selectable 
load-parallel crossover for selective discharging of buffered loads. 

A digital input signal can be used to generate on- and off-durations of any time down to the 
microsecond range for both the power switch and the signal line switches. 

By these features, the TOE 9260 especially is applicable for the following tests according to 
the automotive standards LV 124 and LV 148: 

  LV 124, 

E-10:  Short interruptions 

  LV 124, 

E-13:  Pin interruption 

  LV 148,  E48-09:  Short interruptions 

There are many other industry standards based on LV 124 and LV 148 like VW 80000, 
GS 95024-2, VW 82148, etc. with power and signal line interruptions, where the Electronic 
Switch TOE 9260 can be applied. 

 

Subject of this document is how these tests are carried out by use of the TOE 9260, 
especially under consideration of the following points: 

  Test assembly and wiring of the components 

  Choice of the right switch, i.e. power switch resp. signal line switch of the TOE 9260 

  Reference measurements according to the requirements of the LV 124-2013 

 

 

1.1 

Required test equipment 

Commonly the following components are needed for the test assemblies: 
 

  DUT (Device under Test) 

  Power supply according to the requirements of the Device under Test 

  Electronic Switch TOE 9260 

  4x high current cable (cross section 16 mm²) with connector KBT6AR-N/16-S 

(Manufacturer: Multi-Contact)

 

to connect the power supply and the DUT to the power 

switch of the TOE 9260. 

Note 

Please note, that the connector KBT6AR-N/16-S can only be mounted to a cable by use of 
special tooling, soldering is not possible. 

The use of the following connection cables of length 1,20 m or 0,50 m 
TOE 9260/22 (Red) and TOE 9260/23 (Blue) is recommended: 

Item No. 

Item name 

TOE 9260/22 

0.50 m connection cable with 1 safety socket, red 

TOE 9260/23 

0.50 m connection cable with 1 safety socket, blue 

TOE 9260/24 

1.20 m connection cable with 1 safety socket, red 

TOE 9260/25 

1.20 m connection cable with 1 safety socket, blue 

  Pulse source, e.g. Arbitrary Function Generator TOE 7761, or a DAQ-card with digital 

outputs 

  PC as control device for the function generator or DAQ-card to generate the pulse 

sequences especially for the tests LV 124 E-10 resp. LV 148 E48-09 

  Oscilloscope or fast data logger for evaluation of the measurement results 

 

Summary of Contents for TOE 9260 Series

Page 1: ...al Line Interruptions for Automotive Tests Electronic Switch TOE 9260 9260E AN001_Rev00 doc R TOELLNER ELECTRONIC INSTRUMENTE GMBH Gahlenfeldstrasse 31 58313 Herdecke Germany 49 0 23 30 97 91 91 Fax 4...

Page 2: ...E 10 Short interruptions reference measurements 6 2 3 LV 124 Test E 13 Pin interruption 6 2 3 1 E 13 interruption of signal pins 6 2 3 2 E 13 interruption of signal pins reference measurements 7 2 3 3...

Page 3: ...ice of the right switch i e power switch resp signal line switch of the TOE 9260 Reference measurements according to the requirements of the LV 124 2013 1 1 Required test equipment Commonly the follow...

Page 4: ...ence resistors 1 100 and 1 k of low inductance 5 tolerance to perform reference measurements The use of the following reference resistor kits is recommended Item No Item name TOE 9260 100 Reference Re...

Page 5: ...een the power supply and the IN sockets of the TOE 9260 the internal capacitor at the input of the TOE 9260 should be activated by setting the DIP switch INPUT C to ON So the shortest possible rise an...

Page 6: ...possible 2 3 LV 124 Test E 13 Pin interruption The test E 13 has to be applied to signal pins as well as to ground pins For both situations the test assemblies are explained in the following 2 3 1 E 1...

Page 7: ...istor TOE 9260 104 may be connected to power source and signal line switch by provided suitable cables Due to the 1 A current limit of a signal line switch the measurement with the 1 resistor shall be...

Page 8: ...nce resistor of 1 5 resp 1 k 5 of low parasitic inductance To avoid the addition of too high cable inductivity the reference resistor shall be connected close to the signal line switch by closely spac...

Page 9: ...resistor at a voltage level of 1 V If the DUT is tested with two paralleled signal line switches 2 3 3 the reference measurements have to be carried out with the same configuration of paralleled switc...

Page 10: ...ould be activated by setting the DIP switch INPUT C to ON So the shortest possible rise and fall times of the voltage at the DUT are achieved If an artificial network between the power supply and the...

Reviews: