172-65639M-01 (EF200) 13 Apr 2018
61
8.2.20 Diagnostics
Descriptions of functions: Diagnostics
Actual diagnostics
Displays the current diagnostic event along with the
diagnostic information.
NOTE: If two or more messages occur simultaneously, the
message with the highest priority is shown on the
display.
Displayed: Symbol for diagnostic behavior, diagnostic code
and short message.
Previous
diagnostics
Displays the diagnostic event that occurred prior to the current
diagnostic event along with the diagnostic information.
Displayed: Symbol for diagnostic behavior, diagnostic code
and short message.
Operating time
from restart
Shows the time the device has been in operation since the
last device restart.
Display format: Days (d), hours (h), minutes (m), seconds (s)
Operating time
Indicates how long the device has been in operation.
Display format: Days (d), hours (h), minutes (m), seconds (s)
Diagnostic list
In the Diagnostic list submenu, up to 5 currently pending
diagnostic events can be displayed along with the related
diagnostic information.
NOTE: If more than 5 diagnostic events are pending, the events
with the highest priority are shown on the display.
8.2.21 Event logbook: Diagnostics
→
Event logbook
Descriptions of functions: Event logbook
Filter options
The category of event messages displayed in the event list
can be defined.
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information (I)
Event list
A maximum of 20 event messages can be displayed in
chronological order.
The event history includes entries for:
• Diagnostic events
→
(see 10.3 for details)
• Information events
→
(see 10.4 for details)
In addition to the operation time of its occurrence, each event
is also assigned a symbol that indicates whether the event
has occurred or is ended:
•
Event has occurred
•
Event has ended
8.2.22 Device information: Diagnostics
→
Device information
Descriptions of functions: Device information
Device tag
Enter the name for the measuring point. Max. 32 characters.
Allowed characters: letters, numbers or special characters
(e.g. @, %, /)
Factory setting: EF200
Serial number
Displays the serial number of the measuring device.
Max. 11-digit character string comprising letters and numbers.
Firmware version
Displays the device firmware version installed.
Display format: xx.yy.zz