172-65639M-01 (EF200) 13 Apr 2018
52
100% bargraph
value 1
Enter 100% value for bar graph display. (Signed floating-point
number)
Factory setting: 1 m³/h
Value 2 display
Select the measured value that is shown on the local display.
(see Value 1 display)
Factory setting: Totalizer 1
Value 3 display
Select the measured value that is shown on the local display.
(see Value 1 display)
Factory setting: None
0% bargraph value 3 Enter 0% value for bar graph display. (Signed floating-point
number)
Factory setting: 0
100% bargraph
value 3
Enter 100% value for bar graph display. (Signed floating-point
number)
Factory setting: 0
Value 4 display
Select the measured value that is shown on the local display.
(see Value 1 display)
Factory setting: None
8.2.9 Output conditioning: Setup
→
Output conditioning
Descriptions of functions: Output conditioning
Display damping
Set display reaction time to fluctuations in the measured value.
Setting range: 0.0 to 999.9 seconds
Factory setting: 5.0 seconds
Damping output 1
Set the reaction time of the output signal of the current output
to fluctuations in the measured value.
Setting range: 0.0 to 999.9 seconds
Factory setting: 1 second
Damping output 2
Set the reaction time of the output signal of the frequency
output to fluctuations in the measured value.
Setting range: 0.0 to 999.9 seconds
Factory setting: 1 second
8.2.10 Low flow cut off: Setup
→
Low flow cut off
Descriptions of functions: Low flow cut off
Assign process
variable
Select process variable for low flow cut off.
• Off
• Volume flow
• Corrected volume flow
• Mass flow
• Reynolds number
Factory setting: Off
On value low flow
cutoff
Enter on value for low flow cut off. (Positive floating-point
number)
Factory setting: 0
Off value low flow
cutoff
Enter off value for low flow cut off.
Setting range: 0 to 100.0%
Factory setting: 50%