11.2 OEM menu
Basic settings and limiting values can be changed in this menu.
To make changes in the OEM menu , the system control must previously be unlocked (see
"Unlocking the system").
Calling the OEM menu:
Simultaneous pressing of the Interval-key and the Nonstop-key calls the OEM menu. Following
this, the prompt
"OEM menu Press enter"
appears. When this is confirmed with the Enter-key,
the first menu prompt can be worked on. To simplify changing settings, press the UV-key to
select the number in a value which you want to change. Using the arrow keys now allows a
number from 0 to 9 to be entered at that position.
A press on the menu-key takes you to the next menu prompt.
The OEM menu prompt display shows:
OEM Menu
Press enter
11.2.1 Maximum temperature:
A single press on the menu-key:
The maximum temperature to which the system can be exposed can be set in this menu. When
this temperature is exceeded, the "
max. Temperature"
fault message is triggered. Settings
above 50 °C cause the limiting value to be suppressed and the word
„off“
appears in the
display. It is shown in the 4
th
line of the display.
Basic setting:
50 °C
Setting range:
1 - 50 °C
The display shows:
OEM Menu
max. Temp
50 °C