4.3.2. Display Appearance
Decrease Total for -ve flow:
The meter will not register
negative flow on its outputs unless this box is checked.
For reverse flow the rate will show a “
-
”
sign before the rate
and the total will be reduced accordingly.
One of the transistor outputs (PNP or NPN on Pin 6 or 7)
are configurable to give a logic level when reverse flow is
detected.
Once selected, the totaliser will adjust with the negative flow
and the NPN and PNP outputs will give a pulse output in
proportion to reverse flow seen.
The analog pins will give an output if scaled appropriately.
4.3.3. LCD Options
Display Mode:
Allows user to select from
:
•
Rate
–
The display will show flow rate only.
•
Total
–
The display will show total flow only.
•
Both
–
The display can be cycled from rate to total
using the Left-Hand button on the Atrato or the left
button in the “Flow Details” window.
The Total can be Reset at any time on the display
model by pressing the Right-Hand Button when the
Total is being displayed. Alternatively, a remote
input can be used if either pin 6 or pin 7 has been
utilised.
•
Batch
–
By choosing batch mode the Relay can be
switched and utilized (via external equipment) to
control flow to deliver a fixed-
sized ‘batch’ of fluid.
In batch mode the Left-Hand button is used to Start
or pause the batch, and the Right-Hand button is
used to R
eset or ‘prime’ the system for the next
batch after the current batch has been delivered.
For more detail see Relay section.