TE-5170DV-BL
19
Operations Manual
Note: Calibration orifices should be sent back to Tisch Environmental for calibration on
an annual basis per US EPA Compendium Method IO-2.1 Part 7.3.2
Sampling of Ambient
Air For Total Suspened Particulate Matter (SPM) and PM
10
Using High Volume (HV)
Sampler
1.
Enter the following information in the corresponding cells in the worksheet:
Site Information
Location
The location of the instrument
Date
Current Date
Tech
Technician performing the calibration
Serial #
Serial number of the instrument, Pxxxx
VFC G
Factor
The g-factor of the VFC you are calibrating. This can be found on the
lookup table documentation (first page of this doc) or the sticker
located on the VFC.
Calibration Orifice Information
Make
The make of the orifice, typically Tisch Environmental
Model
The model number of the orifice, typically TE-5028A
Serial #
The Serial number of the calibration orifice you are using
Qa Slope
(m)
The Qa slope of the calibration orifice you are using. This is found
on the calibration documentation provided with the calibration
orifice
Qa Int (b)
The Qa intercept of the calibration orifice you are using. This is
found on the calibration documentation provided with the
calibration orifice
Calibration
Due Date
The date that the calibration of the orifice is due. Orifices should
be calibrated on an annual basis. Call Tisch Environmental at 1-
TSP-AND-PM10 to schedule a calibration.
Ambient Conditions
Temp (Deg
Enter the current ambient temperature at calibration, Ta in
Summary of Contents for TE-5170DV
Page 2: ...TE 5170DV BL 2 Operations Manual TE 5170DV BL TSP VFC...
Page 37: ...TE 5170DV BL 37 Operations Manual Assembly Drawings TE 5003V Filter Holder...
Page 38: ...TE 5170DV BL 38 Operations Manual Calibration Worksheets...
Page 39: ...TE 5170DV BL 39 Operations Manual...
Page 40: ...TE 5170DV BL 40 Operations Manual Calibration Certificate...