MD100_4 10/2014
Operating messages
Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.
A user calibrated method is indicated by an arrow while the test result
is displayed (see "Factory calibration reset").
Error codes
Light absorption too great. Reasons: e.g. dirty optics.
Calibration factor "out of range"
Too much light reaching the detector.
Too much light reaching the detector.
Battery capacity was too low during measurement. Change battery.
SuS:
Factory calibration incorrect / erased
SuS:
User calibration incorrect / erased
Hı
Lo
btLo
GB
Operating messages – Error codes
E 70
E 71
Time Cal
Store Date
RESULT
Cal
E 10 / E 11
E 22
E27 / E28 / E29
E23 / E24 / E25
E 20 / E 21