6
6. Setting Up
• The adjustment knobs located beneath the sensor head
(see diagram “F”) are factory set to “Walk Test Mode”.
Double check they are set as follows;
TIME – Fully anti-clockwise.
DUSK – Fully clockwise.
• Turn the power to the unit ON. The lamp will immediately illuminate
as the unit goes through its “warm-up” period. After approximately
1 minute the lamp will extinguish. This indicates the unit is wired
correctly and the unit is in Test Mode.
• Try to remain outside the detection area during the warm-up period.
• The unit will now operate during daytime as well as at night, illuminating
the lamp for approx. 5 seconds each time. This allows testing to be
carried out to establish whether the sensor is covering the required area.
• Walk around the sensor to establish the detection area.
• The sensor will detect within an approximately 9 metre diameter circle
from the centre of the sensor location with a 3m ceiling. As you cross
a detection “zone” the lamp will illuminate.
• Now stand still until the lamp extinguishes (this should take
approx. 5 seconds). Start moving again after 2 seconds.
As you cross each “zone” the lamp will illuminate.
• Repeat the above, walking at various distances and angles to the unit.
This will help you to confirm the detection pattern.
7. Setting Up for Automatic Operation
• When the walk tests are complete, the unit can be adjusted for
automatic operation.
• The TIME setting controls how long the unit remains illuminated
following activation & after all motion ceases.
• The minimum time (fully anti-clockwise) is approx. 5 seconds, whilst the
maximum time (fully clockwise) is approx. 18 minutes. Set the control
to the desired setting between these limits.