Introduction 

 

1.1

 

Description 

 

The 7194A benchtop pneumatic pressure pump is a dual pressure source which can be 

adjusted and stabilized quickly and easily. With a long pump lever, high output pressures 

can be reached and a high-quality screw press is incorporated to allow fine pressure 

adjustment settings within 0.1 mbar (10 Pa). A pump isolation valve shuts-off the pump from 

the system under pressure to keep the output stable under calibration.   

A built-in gas-liquid trap protects the pump from moisture and dirt.  Any residual liquid in the 

pump will be vented out via the vent port which incorporates a hose connection to allow 

collection of liquids.  Two finger-tight connectors installed on the pump allow the user to 

connect and disconnect devices’ under test, as well as reference gauges quickly and easily. 

The 7194A is simple to operate, with features that make it ideal for calibrating pressure 

transducers, precision pressure gauges, gauge comparison as well as many other pressure 

instruments.  It is particularly suitable for applications where high pressure is required without 

the use of hydraulic pumps or nitrogen cylinders.  

 

 

Summary of Contents for 7194A

Page 1: ...sure Calibration Pump Version 1 5 May 2021 Time Electronics Ltd Unit 5 TON Business Park 2 8 Morley Road Tonbridge Kent TN9 1RA United Kingdom T 44 0 1732 355993 F 44 0 1732 350198 mail timeelectronic...

Page 2: ...Electronics Ltd This also applies to any schematics drawings and diagrams contained herein This manual provides operating and safety instructions for the Time Electronics product To ensure correct op...

Page 3: ...atic Calibration Pump Page 3 of 11 Contents 1 Introduction 4 1 1 Description 4 1 2 Specifications 5 1 3 Options 6 2 Controls 7 2 1 Air Routine 7 3 Operation 8 4 Operating Precautions Troubleshooting 9...

Page 4: ...p the output stable under calibration A built in gas liquid trap protects the pump from moisture and dirt Any residual liquid in the pump will be vented out via the vent port which incorporates a hose...

Page 5: ...ted Pressure Range 0 95 bar vacuum to 200 bar pressure 14 psi vacuum to 3000 psi pressure Over Pressure Range To 235 bar 3500 psi Pressure Setting Resolution 0 1 mbar 10 Pa Material Ram adapters 316 S...

Page 6: ...A Pump Maintenance Kit 2 x Valve pads 2 x Valve cores 2 x Single valve 1 x Torsional spring 1 x Guarded switch 1 x O ring 1 5X2 used for high pressure cylinder 3 x O ring 4X2 used for high pressure cy...

Page 7: ...er 1 4 BSP 1 4 NPT M20 2 Pressure Lever 3 Pressure Vacuum Selector pull out for pressure push in for vacuum 4 Selector Safety Switch 5 Fine Adjust Handles turn clockwise to increase pressure 6 Clean o...

Page 8: ...Time Electronics User Manual 7194A Pneumatic Calibration Pump v1 5 7194A Pneumatic Calibration Pump Page 8 of 11 3 Operation...

Page 9: ...surise device under test Hard to pump pressure lever Check isolating valve is open Check airline for blockages Pressure lever pumps but pressure is not increasing Check release valve is closed Check p...

Page 10: ...ing from action of the user such as misuse operation outside of specification improper maintenance or repair or unauthorized modification Time Electronics total liability is limited to repair or repla...

Page 11: ...ping charges paid to Time Electronics Ltd Unit 5 TON Business Park 2 8 Morley Road Tonbridge Kent TN9 1RA United Kingdom Tel 44 0 1732 355993 Fax 44 0 1732 350198 Email mail timeelectronics co uk Web...