background image

 

                                                                WH-C 

 

          . 

10 

.

 

5. TROUBLESHOOTING 

 

Display 

Probable causes 

Remedy 

Ad

﹉﹉﹉

 

Load cell wires to indicator are 
incorrectly connected, or 
shorted, or opened; or ADC, load 
cell are damaged   

Make sure wires are ok and 
correctly connected. Replace 
load cell or ADC chip, Service 
required. 

Ad

﹍﹍﹍

 

0

﹉﹉﹉﹉

 

Weight reading exceeds Power 
On Zero limit. 

Make sure scale platform is 
empty. 
Perform zero calibration. 

0

﹍﹍﹍﹍

 

Weight reading below Power On 
Zero limit. 

Install platform on scale. 
Perform zero calibration. 

﹉﹉﹉﹉

 

Weight reading exceeds 
overload limit, or The weight 
value cannot be displayed in the 
current unit of measure because 
it exceeds 6 digits. 

Reduce load on scale until 
weight value can be displayed. 
Use a more appropriate unit of 
measure. Contact your local 
dealer or supplier for a new set 
up. 

﹍﹍﹍﹍

 

Weight reading below Under load 
limit. 

Install platform on scale. 
Perform zero calibration 

EEP.E1 

CONFIG or CAL parameters are 
not correctly set 

Contact your local dealer or 
supplier for a new set up 

EEP.E2 

USER parameter is not correctly 
set 

Contact your local dealer or 
supplier for a new set up 

CAP.ER 

Capacity parameters are not 
correct 

Contact your local dealer or 
supplier for a new set up 

CAL.Er 

Calibration error, maybe input 
data or loaded weight is too 
small, too big, unstable, un-linear 

Input correct data, load correct 
weight onto platform, Service 
required 

Not turn on. 

Power cord not plugged in or 
properly connected. Power outlet 
not supplying electricity. Battery 
discharged.   
 
 
Other failure. 

Check power cord connections. 
Make sure power cord is 
plugged into the power outlet. 
Check power source. Replace 
batteries.   
 
Service required. 

Scale can’t set to 

zero or will not zero 

when turned on 

Load on scale exceeds allowable 
limits. 
Load on scale is not stable. 
Load cell damage. 

Remove load on scale. 
Wait for load to become stable. 
 
Service required. 

Empty battery 

symbol/ Lo.bAt is 

shown 

Batteries are discharged. 

Charge batteries   

 

Summary of Contents for TWH-01

Page 1: ...TWH 01 Electronic Physician Scale with BMI Operation Manual Contents subject to change without notice Version 1 0 2019 4 22...

Page 2: ...play Descriptions 4 3 2 Key Description 5 3 3 Indicator Display Character Definitions 6 4 OPERATIONS 7 4 1 Correct Weighing 7 4 2 How to change the unit 7 4 3 Tare function 7 4 4 Weighing with preset...

Page 3: ...vices emit RF and can cause unstable scale readings If your scale ever performs poorly try moving the scale to a different room or location Avoid using in condition of heavy vibration and airflow Read...

Page 4: ...the indicator bracket on the pillar bracket 4 Insert the 2 screws 2 through the holes on the pillar bracket tighten with nuts 18 and flat washers 17 secure the bracket of pillar to the indicator brac...

Page 5: ...rew on the upper bracket 8 9 Locate the height rod 7 onto the fixing points 9 of the upper bracket 8 lower bracket 12 10 Secure the bolt and washer to the bracket on the base 10 11 Tighten the bolt on...

Page 6: ...WH C 4 3 OVERVIEW OF CONTROLS AND FUNCTIONS 3 1 Display Descriptions...

Page 7: ...WH C 5 3 2 Key Description...

Page 8: ...WH C 6 3 3 Indicator Display Character Definitions...

Page 9: ...o get proper result 4 3 Tare function The tare function allows ignoring a fixed weight e g wheelchair that will be deducted after each weighing procedure Proceed as follows Press the ON OFF ZERO with...

Page 10: ...ave one tare weight Entering a new tare weight will automatically replace the old one 4 5 HOLD function The Hold function is being used if you like to hold the results at the display after the weight...

Page 11: ...he patient got a normal weight If he feels well there is no need for a change BMI between 25 and 30 The patient is slightly overweighed to pre obese A reduction of weight is indicated or is necessary...

Page 12: ...on EEP E1 CONFIG or CAL parameters are not correctly set Contact your local dealer or supplier for a new set up EEP E2 USER parameter is not correctly set Contact your local dealer or supplier for a n...

Reviews: