4-20
9) Allow a few moments for stability of the flow indication. This "no flow" condition
will be used for a LOW flow reference. Press [LOW FLOW] to store this point.
Then press <ENTER>.
10) Using the 1.0 liter data obtained in step 2, calculate the following:
1.0 + (Reference Indication) - (3900 Indication)
Press [MID FLOW] and enter this calculated value.
11) Using the 2.0 liter data obtained in step 3, calculate the following:
2.0 + (Reference Indication) - (3900 Indication)
Press [HIGH FLOW] and enter this calculated value.
Count l/min
*Mass Flow
4
.0144
Low
Mid
High
0_
1.0
2.0
-/
/+
OOPS
Count l/min
*Mass Flow
2500
1.000
Low
Mid
High
0
0.98 _
2.0
-/
/+
OOPS
Count l/min
*Mass Flow
5000
2.000
Low
Mid
High
0
0.98
1.97 _
-/
/+
OOPS