background image

THORNTON

Leading Pure Water Analytics

200CR Conductivity/Resistivity Instrument Initial Set-Up 

 

 

200CR Front Panel 

 
The keypad has 6 keys which access specific menus 
as follows:  

measure

 - menus to change measurement modes. 

setpoint

 - menus for programming setpoints. 

relays

 - menus for programming relays. 

outputs

 - menus for programming outputs.  

calibrate

 - menus to perform calibration. 

menus

 - all other menus (cell constants, security, 

 

  averaging, compensation, etc.) 

 
The control keys which are used to make changes 
within a menu are: 

 

OK/Next

 is used to accept a selection and proceed to 

the next menu level. 

 

Up arrow

 is used to scroll up through a list of  

options (& increase numbers). 

 

Down arrow 

is used to scroll down through a list of 

options (& decrease numbers).   

 

Left arrow 

is used to move the cursor to the left 

within a menu. 

 

Right arrow

 is used to move the cursor to the right 

within a menu. 

 
Each digit can be scrolled through the values: 

.

 (decimal point), 0, 1, 2, 3, 4, 5, 6, 7, 8 and 9. The first 

digit of each number can also be set to neg.(-) 

 

Following are the steps necessary to install a 200CR 
instrument and begin operation. 

 
1.  Instrument installation - (Chapter 2) 

 

     The 200CR can be panel, pipe or wall mounted  
     and a sealed IP65 rear cover is optional but is 

required for wall and pipe mounting.  Drill holes in 
the rear cover as needed for conduit or cable grips. 

 
2.  Wiring - (Chapter 2)  
 

Make all necessary electrical connections to the  
instrument.  The wiring  procedure is outlined  on 
the back of this sheet. 

3.  Instrument Calibration - (Chapter 8) 

 

This instrument is factory calibrated within 
specifications and does not require re-calibration. If 
Quality Assurance requirements call for verification, 
follow the procedures in the manual. 

 

4.  Connect sensors to patch cords. 
 
5.  IMPORTANT: Enter cell (sensor) constants for 

resistance and temperature for each channel. 

 Press 

menus

 key then use arrow keys until: 

Edit Sensor Cal  

 Press 

OK/NEXT

 key 

A Cell 

M=.10000_ 

 Select 

(

A Cell, A Temp, B Cell, B Temp

using up and down arrow keys 

 

Shift cursor using the right arrow key to enter 

M

 

, the multiplier. 

 

Shift cursor using the right arrow key to enter 
the precise value of the cell constant found on 
the sensor. 

 

Shift cursor back to the first field and repeat 
the above procedure for each of the other three 
constants.  

 Press the 

OK/NEXT

 key 

Save Changes?

Yes

 

 Press the 

OK/NEXT

 key 

 

6.  Select desired measurements for each sensor 

 Press 

measure 

key 

Channel Primary (A,B) settings: 
For resistivity, ohm-cm  (Auto) - recommended. 
For conductivity, S/cm  (Auto) - recommended. 

By selecting Auto, the instrument will automatically
scale the sensor value to be read by the 
instrument. 

 

Channel Secondary (a,b) settings: 
Secondaries are usually temperatures (

°

F, 

°

C) 

 Press the 

OK/NEXT

 key once all 4  measurement 

selections are made. 

Save Changes?

Yes

 

 Press the 

OK/NEXT

 key to save changes. 

 
7.  Program the analog outputs. (Chapter 7).  Do not 

calibrate analog outputs. 

 

For additional information refer to Manual 84295.  For 
coverage of digital RS232/RS422 communications 
refer to Manual 84364.

 

Reviews: