TROUBLESHOOTING

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 
Question (Q):   

My loss is higher every time I try to re-insert my fiber(s)? 

Answer (A):

     The index matching gel is running out. Every time the fiber is retracted from the 

ULTRA

splice, the gel is    

                          being withdrawn from the glass capillary, thus not leaving enough for proper filling of “gaps”. 
 

Solution (S):

   Use another 

ULTRA

splice, or add more index matching gel (ACA Part # WFN1CCIMG is recommended). 

 

Q:   

My fiber keeps breaking before I get it into the glass capillary? 

A:

   There are a variety of answers to this question: 

 

1)   If using Singlemode Fiber, is a Singlemode (yellow body) 

ULTRA

splice being used? – And if using Multimode    

                    125 micron fiber, is a Multimode (orange body) 

ULTRA

splice being used? 

 

2)  Is the Collet (gray locking nut) adjusted properly to your buffer size(s)? – (Described in section 3) 

 

3)  Is the cleave length of your fiber(s) about 7 millimeters in length? 

 

4)  If working with different buffer sizes (i.e. 900 micron) did you open the gray locking nut too far? – This action 

prevents the collet from guiding the fiber into the glass capillary, and generally the fiber breaks because it has 
nowhere to go, and therefore, it hits the side-wall of the splice body. 

 

5)  Are you treating the 

ULTRA

splice as a copper splice and forcing the fiber? – Especially on the second fiber 

being installed, you have to insert the fiber with care. (As described in section 4). 

 

Q:

   When and Why should I use the Blue Tube? 

A:

   The Blue Tube has no other mechanical purpose except to aid in guiding smaller diameter buffers into the glass  

       capillary. This includes generally anything under 500 micron in buffer size. Some buffers are out of tolerance or larger  
       for the Blue Tube, if this occurs, simply remove the Blue Tube, and try inserting the buffer in that manner (also  
       described in section 3). 
 

Q:   

Do I need to add Index Matching Gel when I get the splice? 

A:   

NO – The 

ULTRA

splice comes pre-loaded with index matching gel, ready to use. 

S:   

The ONLY time that you need to add index matching gel is when the splice is re-used more than four (4) times. The 

reason is because every time the fiber(s) are withdrawn, gel is also being removed, thus gel needs to be replenished. 
 
 

 

There are also a variety of supporting products available for the 

ULTRA

splice, including a Lighted Workstation, 

Holders, and other user friendly products to make the splice easy to install and operate. if you require any further 
information, please contact your local Distributor, or us in any of the following convenient ways: 

 

Thorlabs, Inc.

 

Phone: (973) 579-7227 

Fax: (973) 383-8406 

Email: www.thorlabs.com 

Reviews: