Cascading the interlock of multiple TXP5016 mainframes
Up to five TXP5016 mainframes can be cascaded in order to use a
common interlock.
The interlock opens and the safety function of all cascaded
mainframes is triggered as soon as the current is interrupted. In
other words, an active-open switch should be placed then between
contacts 5 and 9 of the first mainframe.
Note
When cascading two or more TXP5016 interlocks, only one constant
current source must be used!