6.4 IEEE Error Messages
[1103, "Over temperature"]
Reason:
O
ver
T
emperature
P
rotection was tripped. The module is overheated and cannot be
switched on. Wait until the module has cooled down.
[
1104,"Wrong or no sensor"
]
Reason: Attempt to switch on the TEC controller, while no sensor is connected or wrong sensor is
selected.
[
1105,"No calibrating of sensor during TEC on"
]
Reason: The sensor cannot be calibrated while the TEC output is switched on.
[
1106,"Wrong command for this sensor"
]
Reason: The used command is not allowed for the selected sensor, e.g., attempt to set a
thermistor resistance value, while the selected and recognized sensor is an AD590)
[
1107,"No sensor change during TEC on allowed"
]
Reason: The sensor type cannot be changed while the TEC output is switched on.
[
1130,"Command not valid for this module"
]
Reason: A command was entered that is not valid for the present module type (e.g. “CALTR” with
a KRYO module).
6.5 Status Reporting
The TED8000 module provides
DEC
Device Error Condition Register
DEE
Device Error Event Register
EDE
Device Error Event Enable Register
together with
of the PRO8000 mainframe
ESR
Standard event status register
ESE
Standard event Status Enable Register
STB
Status Byte Register
SRE
Service Request Enable Register
to program various service request functions and status reporting.
See also
, section 11.
40
41
50
Summary of Contents for TED8000 Series
Page 1: ...PRO800 PRO8000 4 Series Modular Temperature Controllers TED8000 Operation Manual 2016 ...
Page 2: ...Version Date Item No 3 2 08 Sep 2016 M0009 510 227 Copyright 2016 Thorlabs ...
Page 42: ... 2016 Thorlabs 40 TED8000 Structure of the registers DEC DEE and EDE ...
Page 43: ... 2016 Thorlabs 6 Communication with a PC 41 Structure of the registers ESR ESE STB and SRE ...
Page 59: ......
Page 60: ...www thorlabs com ...