THORLABS SPDMH2 Operation Manual Download Page 11

Rev: 1.1,  23-May-2022

© 2022 Thorlabs

Page  7

SPDMHx Series

3 Operation

·

Focusing  Quality:  For  optimum  timing  resolution  the  light  should  be  focused  to  a  small  spot
(<70 

m

m) in the center of the sensor. Off-center focusing or overfilling the sensor area might

lead to an increased FWHM of the photon timing resolution. This is particularly important for
the free space detectors. 

·

Count Rate: High count rates decrease the timing resolution. Especially at count rates above
1 MHz the FWHM may be double the value compared to low count rates.

Temporal Stability

The temporal stability of the pulse output depends on the count rate. High count rates lead to a
relative shift of the pulse to later times. The total shift may reach 800 ps at count rates above 1
MHz.

Saturation Level

The dead time limits the measurable count rate at high incoming light levels. The count rate at
which the signal does not change significantly by increasing incident photon numbers is called
saturation level. Precautions should be taken to avoid prolonged excessive light levels that may
damage the SPDMHx Series detector.

Correction Factor 

Every  SPDMHx  Series  detector  has  an  inherent  dead  time  of  approx.  43  ns  after  detecting  a
photon. The  dead  time  is  also  noted  in  the  included  production  report.  During  this  dead  time,
the SPDMHx Series detector is "blind" and can not detect further photons. As a consequence,
the measured count rate is lower than the true incident photon rate.

The photon rate can be calculated from the measured counting rate as follows: 

R

R

1-R

T

measured

photon

measured

D

=

g

where: 

R

photon

   : actual incident photon rate

R

measured  

: measured count rate

T

D  

: detector dead time

The correction factor can be used to correct the non-linearity especially at high light levels. 

The following plot shows the effect of the dead time as the measured count rate does not pro-
portionally increase with the real photon rate for high count rates due to the dead time effect.
The correction factor is required to receive the real photon rate. 

Summary of Contents for SPDMH2

Page 1: ...Single Photon Detectors SPDMH2 SPDMH3 SPDMH2F SPDMH3F Operation Manual 2022...

Page 2: ...Version Date 1 1 23 May 2022 Copyright 2022 Thorlabs...

Page 3: ...ting Function and TTL Output 6 3 4 2 Optimizing Performance 4 Maintenance and Service 9 5 Appendix 10 10 5 1 Technical Data 11 5 2 Performance Plots 11 5 2 1 Photon Detection Efficiency 12 5 3 Dimensi...

Page 4: ...king forward to hearing from you Thorlabs GmbH Warning Sections marked by this symbol explain dangers that might result in personal injury or death Always read the associated information carefully bef...

Page 5: ...de op tical fiber with an FC connector item s SPDMH2F or SPDMH3F Applications range from quantum technologies and cryptography to particle sizing fluorescence analysis LIDAR and spectroscopy Attention...

Page 6: ...er appears to be damaged keep it until you have inspected the contents for completeness and tested the SPDMHx Series mechanically and electrically Verify that you have received the following items wit...

Page 7: ...he components of the SPDMH2 and SPDMH3 free space detectors are labeled in the image of the SPDMH2 SPDMH2 and SPDMH3 components look identical SPDMH2F and SPDMH3F The components of the SPDMH2F and SPD...

Page 8: ...weight of the detector into account when this mounting style is used Base Plate Mounting The base plate of the SPDMHx Series detector can be fastened on a bread board using the CL4 table clamps Alter...

Page 9: ...is therefore recommended Please see the Mounting section for more information Ensure that background light does not reach the photosensitive area This can be achieved by mounting lens tubes onto the C...

Page 10: ...window as gating can deselect longer time periods with no signal Also applications with very weak signal and high background light profit from the gating in that the background signal within periods...

Page 11: ...ntly by increasing incident photon numbers is called saturation level Precautions should be taken to avoid prolonged excessive light levels that may damage the SPDMHx Series detector Correction Factor...

Page 12: ...to very low light levels The measured photon rate de creases with increasing optical power Thus at high optical power the measured photon rate will differ from the real photon rate The following grap...

Page 13: ...s is not water resistant Attention To avoid damage to the instrument do not expose it to spray liquids or solvents The unit does not need a regular maintenance by the user It does not contain any modu...

Page 14: ...n 0 5 V 2 4 V Trigger Input Response Time Closing Signal Opening Signal 15 ns Typ to 20 ns Max 60 ns Typ to 65 ns Max Afterpulsing Probability 0 2 Typ Delay between Photon Impact and TTL Pulse 30 ns T...

Page 15: ...GRIN lens fo cuses the light onto a spot of 70 mm diameter in the center of the detector 2 Specifications are valid for modules without FC connector 3 Default in the absence of a TTL signal is 2 4 V a...

Page 16: ...Page 12 MTN028160 D02 SPDMHx Series 5 Appendix 5 3 Dimensions SPDMH2 SPDMH2 and SPDMH3 outer dimensions are identical...

Page 17: ...Rev 1 1 23 May 2022 2022 Thorlabs Page 13 SPDMHx Series 5 Appendix SPDMH2F SPDMH2F and SPDMH3F outer dimensions are identical...

Page 18: ...serts If necessary ask for replacement packaging Refer servicing to qualified personnel Changes to this device cannot be made nor may components not supplied by Thorlabs be used without written consen...

Page 19: ...Rev 1 1 23 May 2022 2022 Thorlabs Page 15 SPDMHx Series 5 Appendix 5 5 Certifications and Compliances...

Page 20: ...starting with the date of shipment in accordance with and subject to the terms and conditions set forth in Thorlabs General Terms and Conditions of Sale which can be found at General Terms and Conditi...

Page 21: ...of Life Policy WEEE Thorlabs verifies our compliance with the WEEE Waste Electrical and Electronic Equipment direct ive of the European Community and the corresponding national laws Accordingly all e...

Page 22: ...www thorlabs com...

Reviews: