![THORLABS SPCM A Series Operation Manual Download Page 38](http://html.mh-extra.com/html/thorlabs/spcm-a-series/spcm-a-series_operation-manual_1110712038.webp)
36
SPCMxxA
© 2019 Thorlabs
20. Get Fan State
SYSTem:FAN[:STATe]?
Returns actual fan state:
0 - fan off
1 - fan on
21. Set LED State
SYSTem:LED[:STATe] {OFF | 0 | ON | 1}
with values
OFF | 0 - led off
ON | 1 - led on
Remark: LED is set on by default. Starting a measurement switches LED off (to avoid influence
of the measurement) and stopping a measurement switches LED on.
22. Get LED State
SYSTem:LED[:STATe]?
Returns actual led state:
0 - led off
1 - led on
Remark: LED is set on by default. Starting a measurement switches LED off (to avoid influence
of the measurement) and stopping a measurement switches LED on.
23. Set DFU State
A command to prepare device for firmware update
STATus:DFU <value>
with values
0 - DFU off
1 - DFU on
2 - DFU once (disappears after device power cycle)
Remark
: Setting DFU state to 1 or 2 results in enumeration of the USB device - the connection
will be lost. Do not power off the device until firmware update is complete. Use Thorlabs DFU
Wizard downloadable from the Thorlabs webside for updating the firmware of the SPCM.
Hint
: Use “DFU once” for a normal firmware update to avoid that a second USB driver (for
DFU) is needed to be installed if the Single Photon Counter is connected to another USB port.
Especially moving the SPC to another PC may be problematic because the needed DFU driver
may not exist on this PC (it comes with Thorlabs DFU Wizard).
Summary of Contents for SPCM A Series
Page 1: ...Single Photon Counting Module SPCMxxA Operation Manual 2019 ...
Page 2: ...Version Date 1 2 27 May 2019 ...
Page 42: ...40 SPCMxxA 2019 Thorlabs ...
Page 43: ...5 Maintenance and Service 2019 Thorlabs 41 ...
Page 54: ...52 SPCMxxA 2019 Thorlabs 6 5 Dimensions Dimensions of the SPCMxxA imperial version ...
Page 55: ...6 Appendix 2019 Thorlabs 53 Dimensions of the SPCMxxA M metric version ...
Page 61: ...www thorlabs com ...