![THORLABS SA201-EC Operating Manual Download Page 14](http://html.mh-extra.com/html/thorlabs/sa201-ec/sa201-ec_operating-manual_1110778014.webp)
Spectrum Analyzer Controller
Chapter 4: Description
Page 10
6679-D02
NOTE
The PDAVJ5 that is recommended for use along with the SA200-30C should not
be connected to this input. Connecting to the amplifier will cause damage to the
instrument.
4.14. PD Amplifier Output BNC
This BNC is the amplifier output and may be connected directly to an oscilloscope
to view the cavity spectrum. The amplifier gain will be set using the front panel
‘DETECTOR’ control knob. The amplifier output includes a 50 Ω series resistor to
minimize noise when operating with a 50
Ω coax cable. For best results, a 50 Ω
load resistor is recommended at the oscilloscope. Note, the amplifier gain will be
halved with a 50 Ω load connected.
4.15. Voltage Selector Switch
The voltage selector switch allows the user to select the input line voltage. The
factory default setting is 100/115 VAC as shown in Figure 4. To operate at 230
VAC, this switch will have to be moved to the 230 V position. The line fuse will also
need to be changed to properly protect the unit. See sections 5.2 below for detailed
instructions.
Summary of Contents for SA201-EC
Page 1: ...SA201 EC Spectrum Analyzer Controller Operating Manual ...
Page 2: ......
Page 27: ......
Page 28: ...www thorlabs com ...