7.3. Fuse
Replacement
It may be necessary to replace an open fuse. To do this you must perform the following procedure.
Remove the AC power cord if it is connected to the unit.
Locate the fuse tray directly below the AC power cord connection on the rear panel of the unit.
Carefully use a flat blade screwdriver to open the fuse tray.
Remove the existing fuse and install the appropriate fuse.
Specification
Value
Manufacturer
Littlefuse
Manufacturer part number
218.500 HXP
Thorlabs item number
4300-GMC-250-500MA
Description
500 mA, 5X20mm, Time-Lag, 218 Series
Push the fuse tray back into place making sure that it snaps and seats correctly.