4) Connect the foot pump to the mount on one of the legs and inflate to about 40 p.s.i. – see Fig. 3.4.
5) Inflate the leg diagonally opposite to 40 p.s.i.
6) Repeat steps 4) and 5) for the two remaining legs.
7) Populate the breadboard.
8) Further inflate the mounts until the breadboard is level and the correct height is obtained – see Fig. 3.5.
9) Fit the dust caps.
Fig. 3.4 Inflating passive anti-vibration mounts
Fig. 3.5 Correct inflation of passive mounts
Caution
During item 8):
it may be necessary to inflate each mount to a different pressure due to the loading of the breadboard.
do not exceed 72 p.s.i.
65 mm (2.5")
65 mm (2.5")
under inflated
over inflated
correctly inflated
standard load