Si Amplified Fixed Gain Detector
Page 8
13054-D02
Chapter 6 Specifications
All performance specifications are typical, performed at 25 °C ambient
temperature, and assume a 50
Ω
load, unless stated otherwise.
Electrical Specifications
Detector
- Si
PIN
Active Area
-
Ø1 mm (0.8 mm
2
)
Wavelength Range
2
λ
200 to 1100 nm
Peak Wavelength
λ
p
730 nm (Typ)
Peak Response
(
λ
p
)
0.44 A/W (Typ)
Small Signal Bandwidth
- 150
MHz
NEP (
λ
p
)
W/
√
Hz
3.5 x 10
-11
Noise (RMS)
- 1.5
mVrms
Max Ouput Current
I
OUT
100
mA
Dark Offset
- ±10
mV
Load Impedance
- 50
Ω
to Hi-Z
Transimpedance Gain
-
5 x 10
3
V/A (50
Ω
)
1 x 10
4
V/A (Hi-Z)
Output Voltage
V
OUT
0 to 5 V (50
Ω
)
0 to 10 V (Hi-Z)
General
On/Off Switch
Slide
Output
BNC (DC Coupled)
Package Size
2.80" x 1.90" x 0.83"
(70 mm x 48 mm x 21 mm)
PD Surface Depth
0.16" (4.1 mm)
Weight, Detector Only
0.15 lbs
Accessories
SM1T1 Coupler
SM1RR Retainer Ring
Operating Temp
10 to 50 °C
Storage Temp
-25 to 70 °C
AC Power Supply
AC – DC Converter
Input Power
3
31 W
100 – 200 VAC (50 to 60Hz)
220 – 240 VAC (50 to 60 Hz)
2
When long-term UV light is applied, the product specifications may degrade. For example,
the product’s UV response may decrease and the dark current may increase. The degree
to which the specifications may degrade is based upon factors such as the irradiation
level, intensity, and usage time.
3
Although the power supply is rated for 31 W the PDA10A actual usage is <5 W over the
full operating range.