OCTG Series Scanner
Chapter 2: Introduction
Page 2
MTN004419-D02
Chapter 2
Introduction
2.1.
Safety
Please read this manual carefully before operating the OCTG standard scanner. Please also read any
manuals for the systems being connected to the OCTG.
All statements regarding safety and technical specifications will only apply when the unit is operated correctly.
ATTENTION
This equipment is intended for laboratory use only and is not certified for medical
applications, including, but not limited to, life support situations.
WARRANTY WARNING
There are sensitive electronic and optical parts in the OCTG.
Any modification or servicing of this system by unqualified personnel renders Thorlabs free of
any liability.
Any modification or the galvanometer scanners or the camera may cause loss of the factory
optical alignment.
This device can only be returned for service when it is packed into the complete original
packaging, including all foam packing inserts. Please contact Thorlabs
’ OCT support
(see Chapter 11) for replacement packaging if the original packaging has been lost.
LASER RADIATION WARNING
When a light source (e.g. SLD, laser) is being coupled into the OCTG, please observe the
appropriate laser safety precautions for your own protection. The appropriate laser safety
precautions depend on the light source coupled into the OCTG.
During normal operations, laser light will be present within the scanner and will also be
emitted from the scanner. Also laser light may be emitted from unexpected locations, such as
if the fiber has been disconnected from the body or if the reference arm has been
disconnected.
In addition, the OCTG is an optical system that can influence the divergence of the beam. This
can cause a change of laser class of the light source, especially if the OCTG is used without
an objective.
Always turn off the light source before changing or adjusting the OCTG configuration or
accessories as the objective, lens kit, reference arm adapter, or sample z-spacer. For Thorlabs
OCT base units, turn off the OCT base unit main power to turn off the light source.